FDA Device Recalls

Recalls /

#67904

Product

CORDIS "Fire Star" 2.00 x 30, Dual Marker Band, Dilatation Catheter, Catalog # 8013020D, Distributed by Cordis Corporation, Miami Lakes, FL 33014

FDA product code
LOXCatheters, Transluminal Coronary Angioplasty, Percutaneous
Device class
Class 2
Medical specialty
Cardiovascular
PMA numbers
P880003
Affected lot / code info
Lot #s: 13206013 13214559 13223384 13224192 13244440 13244444 13264103 13270373 13278344 13281297 13283826 13292701 13298789 and 13303768

Why it was recalled

Slow Deflation or No Deflation

Root cause (FDA determination)

Process control

Action the firm took

Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.

Recalling firm

Firm
Cordis Corporation
Address
14201 NW 60th Avenue, Miami Lakes, Florida 33014-2802

Distribution

Distribution pattern
Worldwide.

Timeline

Recall initiated
2008-01-14
Posted by FDA
2008-02-06
Terminated
2009-10-20
Status