Recalls / —
—#67967
Product
Medtronic IsoMed Implantable Constant-Flow Infusion Pump, model 8472-20. The contents of the inner package have been sterilized by ethylene oxide gas. Reservoir size: 20 mL. Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA. The implantable Medtronic IsoMed Constant-Flow Infusion Pump is part of the IsoMed Constant-Flow Infusion System designed to contain and administer parenteral drugs to a specific site. The components of the IsoMed Constant-Flow Infusion System include the pump, Medtronic catheter, catheter accessories, Medtronic Refill kit, and Medtronic catheter access port kit. IsoMed pumps are implantable devices that store and dispense drugs according to a constant flow rate set during the manufacturing process.
- FDA product code
- LKK — Pump, Infusion, Implanted, Programmable
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P860004, P990034
- Affected lot / code info
- all serial numbers
Why it was recalled
Device/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. On January 16, 2008, Medtronic sent a letter to doctors who implant these d
Root cause (FDA determination)
Device Design
Action the firm took
An Urgent Medial Device Correction letter was sent January 16, 2008, to Health Care Professionals. The letter describes the incidences, symptoms and recommendations for patient management. Excerpts from the approved Medtronic Professional Labeling are also included with the letter. The firm has requested unreported inflammatory mass in a patient with a Medtroinc device to the firm and to the FDA MedWatch Program by phone at 1-800-FDA-1088. Additional assistance may be obtained by contacting Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Recalling firm
- Firm
- Medtronic Neuromodulation
- Address
- 800 53rd Ave NE PO Box 1250, Minneapolis, Minnesota 55440-1250
Distribution
- Distribution pattern
- Worldwide Distribution -- USA including states of Washington D.C., Puerto Rico, and countries of Aruba, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Faroe Islands, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Mexico, Netherlands, Netherlands Antilles, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vatican City State.
Timeline
- Recall initiated
- 2008-01-16
- Posted by FDA
- 2008-03-22
- Terminated
- 2011-06-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #67967. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.