Product
Siemens Magnetom Sonata syngo, Model Number: 7106425, Siemens Medical Solutions USA, Malvern PA, 19355
- FDA product code
- LNH — System, Nuclear Magnetic Resonance Imaging
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K971684
- Affected lot / code info
- Serial Number: 21936
Why it was recalled
Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Root cause (FDA determination)
Device Design
Action the firm took
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 51 Valley Stream Pkwy, Malvern, Pennsylvania 19355-1406
Distribution
- Distribution pattern
- Nationwide Distribution
Timeline
- Recall initiated
- 2007-12-01
- Posted by FDA
- 2008-03-25
- Terminated
- 2012-10-01
- Status
- —