Recalls / —
—#68299
Product
GE Healthcare Precision RXi 23A/ 32A Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, 65 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of exams for various clinical applications with ether 23 cm image intensifier with 512x512 Image Chain or 32 cm Image intensifier with either 512x512 Image Chain or 1Kx 1K Image Chain.
- FDA product code
- JAA — System, X-Ray, Fluoroscopic, Image-Intensified
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K041605
- Affected lot / code info
- 101 103 117 121 128 132 133 135 136 152 156 160 162 164 173 178 182 184 190 194 196 201 202 209 221 228 232 235 238 240 246 250 251 252 254 259 260 261 263 264 267 268 273 275 276 281 283 289 295 297 300 302 310 322 330 331 333 342 343 347 348 351 353 354 365 367
Why it was recalled
GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .
Root cause (FDA determination)
Device Design
Action the firm took
Two letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect and correct the clearance between the console joystick and its mounting location. If your facility has a Precision RXi or Precision e system manufactured before March 2007, a GE Healthcare Field Engineer will schedule a field visit to install the added joystick protection. This upgrade will be performed at no charge. If you have any questions/concerns regarding this issue, please contact your local GE Healthcare sales or service representative.
Recalling firm
- Firm
- GE Healthcare
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- (USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE SI SK SY TN TR UA YU ZA
Timeline
- Recall initiated
- 2007-12-18
- Posted by FDA
- 2008-08-24
- Terminated
- 2012-08-09
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #68299. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.