Product
GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.
- FDA product code
- KPR — System, X-Ray, Stationary
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K993090
- Affected lot / code info
- 00000010838HL1, 00000011061HL9, 00000011067HL6, 00000011185HL6, 00000012500HL5, 00000012667HL2, 00000013129HL2, 00000013372HL8, 00000013385HL0, 00000013543HL4, 00000013557HL4, 00000013800HL8, 00000014099HL6, 00000014104HL4, 00000014319HL8, 00000014327HL1, 00000014667HL0, 00000014677HL9, 00000015076HL3, 00000015383HL3, 00000015386HL6, 00000015400HL5, 00000015738HL8, 00000015744HL6, 00000015745HL3, 00000015750HL3, 00000015760HL2, 00000016174HL5, 00000016383HL2, 00000016386HL5, 00000016696HL7, 00000016707HL2, 00000016708HL0, 00000016709HL8, 00000016930HL0, 00000016954HL0, 00000016955HL7, 00000017119HL9, 00000017130HL6, 00000017131HL4, 00000017476HL3, 00000017477HL1, 00000017478HL9, 00000017480HL5, 00000017490HL4, 00000017492HL0, 00000017493HL8, 00000017599HL2, 00000017604HL0, 00000017608HL1, 00000017610HL7, 00000017613HL1, 00000017616HL4, 00000017788HL1, 00000017801HL2, 00000018476HL2, 00000018479HL6, 00000018483HL8, 00000019041HL3, 00000019048HL8, 00000019095HL9, 00000019098HL3, 00000019104HL9, 00000019110HL6, 00000019241HL9, 00000019244HL3, 00000019413HL4, 00000019419HL1, 00000019489HL4, 00000019496HL9, 00000019499HL3, 00000019501HL6, 00000019502HL4, 00000019669HL1, 00000019801HL0, 00000019886HL1, 00000019894HL5, 00000019959HL6, 00000020091HL5, 00000020092HL3, 00000020239HL0, 00000020243HL2, 00000020374HL5, 00000020967HL6, 00000021064HL1, 00000021197HL9, 00000021201HL9, 00000021204HL3, 00000021343HL9, 00000021344HL7, 00000021349HL6, 00000021552HL5, 00000021911HL3, 00000021921HL2, 00000021938HL6, 00000021953HL5, 00000022033HL5, 00000022333HL9, 00000022337HL0, 00000022412HL1, 00000022477HL4, 00000022600HL1, 00000022795HL9, 00000022931HL0, 00000023116HL7, 00000023122HL5, 00000023126HL6, 00000023412HL0, 00000023415HL3, 00000024386HL5, 00000024483HL0, 00000024588HL6, 00000024939HL1, 00000024940HL9, 00000025426HL8, 00000025475HL5, 00000025476HL3, 00000025477HL1, 00000025572HL9, 00000027104HL9, 00000027867HL1, 00000028057HL8, 00000028453HL9, 00000028524HL7, 00000028845HL6, 00000029098HL1, 00000029145HL0, 00000029502HL2, 00000029550HL1, 00000029634HL3, 00000029640HL0, 00000029969HL3, 00000030108HL5, 00000030113HL5, 00000030229HL9, 00000030232HL3, 00000030236HL4, 00000030633HL2, 00000030848HL6, 00000030920HL3, 00000030921HL1, 00000030976HL5, 00000031172HL0, 00000031175HL3, 00000031565HL5, 00000031702HL4, 00000031811HL3, 00000031928HL5, 00000031929HL3, 00000031930HL1, 00000032309HL7, 00000032311HL3, 00000032317HL0, 00000032319HL6, 00000032322HL0, 00000032850HL0, 00000032912HL8, 00000032916HL9, 00000032919HL3, 00000032920HL1, 00000033330HL2, 00000033388HL0, 00000033391HL4, 00000033408HL6, 00000033437HL5, 00000033509HL1, 00000033679HL2, 00000033681HL8, 00000033682HL6, 00000033727HL9, 00000033731HL1, 00000033884HL8, 00000033970HL5, 00000033971HL3, 00000033983HL8, 00000033984HL6, 00000034218HL8, 00000034420HL0, 00000034469HL7, 00000034470HL5, 00000034520HL7, 00000034644HL5, 00000034694HL0, 00000034762HL5, 00000034764HL1, 00000034913HL4, 00000034938HL1, 00000035052HL0, 00000035101HL5, 00000035159HL3, 00000035191HL6, 00000035244HL3, 00000035246HL8, 00000035247HL6, 00000035329HL2, 00000035445HL6, 00000035544HL6, 00000035547HL9, 00000035632HL9, 00000035641HL0, 00000036038HL8, 00000036068HL5, 00000036071HL9, 00000036100HL6, 00000036126HL1, 00000036128HL7, 00000036265HL7, 00000036362HL2, 00000036363HL0, 00000036431HL5, 00000036624HL5, 00000036625HL2, 00000036674HL0, 00000036698HL9, 00000036702HL9, 00000036882HL9, 00000036983HL5, 00000037104HL7, 00000037123HL7, 00000037351HL4, 00000037352HL2, 00000037353HL0, 00000037396HL9, 00000037596HL4, 00000037919HL8, 00000038423HL0, 00000038771HL2, 00000038772HL0, 00000039125HL0, 00000039581HL4, 00000039612HL7, 00000040366HL7, 00000040493HL9, 00000040728HL8, 00000040865HL8, 00000041290HL8, 00000041291HL6, 00000041511HL7, 00000041768HL3, 00000042367HL3, 00000042649HL4, 00000042650HL2, 00000042651HL0, 00000042701HL3, 00000042880HL5, 00000043029HL8, 00000043184HL1, 00000043265HL8, 00000043665HL9, 00000043727HL7.00000043747HL5, 00000043751HL7, 00000044548HL6, 00000045475HL1, 00000045651HL7, 00000015925HL1, 00000019969HL5, 00000024476HL4, 00000027957HL0, 00000030025HL1, 00000030028HL5, 00000032711HL4, 00000035548HL7, 00000036428HL1, 00000036883HL7, 00000020371HL1, 00000021582HL2, 00000031922HL8, 00000033316HL1, 00000042706HL2, 00000020735HL7, 00000022513HL6, 00000022926HL0, 00000024376HL6, 00000031909HL5, 00000040521HL7, 00000041017HL5, 00000011544HL4, 00000014336HL2, 00000016171HL1, 00000016545HL6, 00000017123HL1, 00000017483HL9, 00000023189HL4, 00000028833HL2, 00000030875HL9, 00000030924HL5, 00000031916HL0, 00000031918HL6, 00000032909HL4, 00000033730HL3, 00000033974HL7, 00000034268HL3, 00000036165HL9, 00000042001HL8, 00000016177HL8, 00000031554HL9, 00000036626HL0, 00000013383HL5, 00000014323HL0, 00000019485HL2, 00000020232HL5, 00000010836HL5, 00000019679HL0, 00000023546HL5, 00000023547HL3, 00000030926HL0, 00000041996HL0, 00000042388HL9, 00000042537HL1, 00000011063HL5, 00000015391HL6, 00000017138HL9, 00000021058HL3, 00000030918HL7, 00000040086HL1, 00000010328HL3, 00000013375HL1, 00000014335HL4, 00000015065HL6, 00000015081HL3, 00000015747HL9, 00000016371HL7, 00000016701HL5, 00000017125HL6, 00000017481HL3, 00000019482HL9, 00000019486HL0, 00000019678HL2, 00000019680HL8, 00000019788HL9, 00000020081HL6, 00000020224HL2, 00000021639HL0, 00000022032HL7, 00000023118HL3, 00000023181HL1, 00000024432HL7, 00000029259HL9, 00000031908HL7, 00000031914HL5, 00000032231HL3, 00000032897HL1, 00000033724HL6, 00000033725HL3, 00000033968HL9, 00000034260HL0, 00000034459HL8, 00000034462HL2, 00000034783HL1, 00000035377HL1, 00000035378HL9, 00000036563HL5, 00000037254HL0, 00000038190HL5, 00000016175HL2, 00000019042HL1, 00000013796HL8, 00000011182HL3, 00000032313HL9, 00000012475HL0, 00000020222HL6, 00000029743HL2, 00000033726HL1, 00000020732HL4, 00000020737HL3, 00000012674HL8, 00000021551HL7, 00000031689HL3, 00000010651HL8, 00000015757HL8, 00000022520HL1, 00000015082HL1, 00000015153HL0, 00000021644HL0, 00000013373HL6, 00000015158HL9, 00000022606HL8, 00000022607HL6, 00000024489HL2, 00000024590HL2, 00000024671HL0, 00000024689HL2, 00000030446HL9, 00000030460HL0, 00000030927HL8, 00000036429HL9, 00000015387HL4, 00000039893HL3, 00000039895HL8, 00000039962HL6, 00000040002HL8, 00000040072HL1, 00000040164HL6, 00000040167HL9, 00000022402HL2, 00000036266HL5, 00000036914HL0, 00000037252HL4, 00000013564HL0.
Why it was recalled
The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
Root cause (FDA determination)
Process control
Action the firm took
GE issued a Field Modification Instruction to their field service representatives, released in December 2007, with instructions to replace the labels on the console with the appropriate labeling. Consignees will be visited by GE field service representatives beginning December 2007. All consignees will have the required labels affixed to their units by July 31, 2008.
Recalling firm
- Firm
- GE Healthcare
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide Distribution, including USA, Canada, Mexico, France, Hong Kong, Italy, Korea, China, Poland, UK, New Zealand, Saudi Arabia, Taiwan, Turkey, and Australia.
Timeline
- Recall initiated
- 2007-11-01
- Posted by FDA
- 2008-06-10
- Terminated
- 2011-12-18
- Status
- —