GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.

FDA product code
KPRSystem, X-Ray, Stationary
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K993090
Affected lot / code info
00000010838HL1, 00000011061HL9, 00000011067HL6, 00000011185HL6, 00000012500HL5, 00000012667HL2, 00000013129HL2, 00000013372HL8, 00000013385HL0, 00000013543HL4, 00000013557HL4, 00000013800HL8, 00000014099HL6, 00000014104HL4, 00000014319HL8, 00000014327HL1, 00000014667HL0, 00000014677HL9, 00000015076HL3, 00000015383HL3, 00000015386HL6, 00000015400HL5, 00000015738HL8, 00000015744HL6, 00000015745HL3, 00000015750HL3, 00000015760HL2, 00000016174HL5, 00000016383HL2, 00000016386HL5, 00000016696HL7, 00000016707HL2, 00000016708HL0, 00000016709HL8, 00000016930HL0, 00000016954HL0, 00000016955HL7, 00000017119HL9, 00000017130HL6, 00000017131HL4, 00000017476HL3, 00000017477HL1, 00000017478HL9, 00000017480HL5, 00000017490HL4, 00000017492HL0, 00000017493HL8, 00000017599HL2, 00000017604HL0, 00000017608HL1, 00000017610HL7, 00000017613HL1, 00000017616HL4, 00000017788HL1, 00000017801HL2, 00000018476HL2, 00000018479HL6, 00000018483HL8, 00000019041HL3, 00000019048HL8, 00000019095HL9, 00000019098HL3, 00000019104HL9, 00000019110HL6, 00000019241HL9, 00000019244HL3, 00000019413HL4, 00000019419HL1, 00000019489HL4, 00000019496HL9, 00000019499HL3, 00000019501HL6, 00000019502HL4, 00000019669HL1, 00000019801HL0, 00000019886HL1, 00000019894HL5, 00000019959HL6, 00000020091HL5, 00000020092HL3, 00000020239HL0, 00000020243HL2, 00000020374HL5, 00000020967HL6, 00000021064HL1, 00000021197HL9, 00000021201HL9, 00000021204HL3, 00000021343HL9, 00000021344HL7, 00000021349HL6, 00000021552HL5, 00000021911HL3, 00000021921HL2, 00000021938HL6, 00000021953HL5, 00000022033HL5, 00000022333HL9, 00000022337HL0, 00000022412HL1, 00000022477HL4, 00000022600HL1, 00000022795HL9, 00000022931HL0, 00000023116HL7, 00000023122HL5, 00000023126HL6, 00000023412HL0, 00000023415HL3, 00000024386HL5, 00000024483HL0, 00000024588HL6, 00000024939HL1, 00000024940HL9, 00000025426HL8, 00000025475HL5, 00000025476HL3, 00000025477HL1, 00000025572HL9, 00000027104HL9, 00000027867HL1, 00000028057HL8, 00000028453HL9, 00000028524HL7, 00000028845HL6, 00000029098HL1, 00000029145HL0, 00000029502HL2, 00000029550HL1, 00000029634HL3, 00000029640HL0, 00000029969HL3, 00000030108HL5, 00000030113HL5, 00000030229HL9, 00000030232HL3, 00000030236HL4, 00000030633HL2, 00000030848HL6, 00000030920HL3, 00000030921HL1, 00000030976HL5, 00000031172HL0, 00000031175HL3, 00000031565HL5, 00000031702HL4, 00000031811HL3, 00000031928HL5, 00000031929HL3, 00000031930HL1, 00000032309HL7, 00000032311HL3, 00000032317HL0, 00000032319HL6, 00000032322HL0, 00000032850HL0, 00000032912HL8, 00000032916HL9, 00000032919HL3, 00000032920HL1, 00000033330HL2, 00000033388HL0, 00000033391HL4, 00000033408HL6, 00000033437HL5, 00000033509HL1, 00000033679HL2, 00000033681HL8, 00000033682HL6, 00000033727HL9, 00000033731HL1, 00000033884HL8, 00000033970HL5, 00000033971HL3, 00000033983HL8, 00000033984HL6, 00000034218HL8, 00000034420HL0, 00000034469HL7, 00000034470HL5, 00000034520HL7, 00000034644HL5, 00000034694HL0, 00000034762HL5, 00000034764HL1, 00000034913HL4, 00000034938HL1, 00000035052HL0, 00000035101HL5, 00000035159HL3, 00000035191HL6, 00000035244HL3, 00000035246HL8, 00000035247HL6, 00000035329HL2, 00000035445HL6, 00000035544HL6, 00000035547HL9, 00000035632HL9, 00000035641HL0, 00000036038HL8, 00000036068HL5, 00000036071HL9, 00000036100HL6, 00000036126HL1, 00000036128HL7, 00000036265HL7, 00000036362HL2, 00000036363HL0, 00000036431HL5, 00000036624HL5, 00000036625HL2, 00000036674HL0, 00000036698HL9, 00000036702HL9, 00000036882HL9, 00000036983HL5, 00000037104HL7, 00000037123HL7, 00000037351HL4, 00000037352HL2, 00000037353HL0, 00000037396HL9, 00000037596HL4, 00000037919HL8, 00000038423HL0, 00000038771HL2, 00000038772HL0, 00000039125HL0, 00000039581HL4, 00000039612HL7, 00000040366HL7, 00000040493HL9, 00000040728HL8, 00000040865HL8, 00000041290HL8, 00000041291HL6, 00000041511HL7, 00000041768HL3, 00000042367HL3, 00000042649HL4, 00000042650HL2, 00000042651HL0, 00000042701HL3, 00000042880HL5, 00000043029HL8, 00000043184HL1, 00000043265HL8, 00000043665HL9, 00000043727HL7.00000043747HL5, 00000043751HL7, 00000044548HL6, 00000045475HL1, 00000045651HL7, 00000015925HL1, 00000019969HL5, 00000024476HL4, 00000027957HL0, 00000030025HL1, 00000030028HL5, 00000032711HL4, 00000035548HL7, 00000036428HL1, 00000036883HL7, 00000020371HL1, 00000021582HL2, 00000031922HL8, 00000033316HL1, 00000042706HL2, 00000020735HL7, 00000022513HL6, 00000022926HL0, 00000024376HL6, 00000031909HL5, 00000040521HL7, 00000041017HL5, 00000011544HL4, 00000014336HL2, 00000016171HL1, 00000016545HL6, 00000017123HL1, 00000017483HL9, 00000023189HL4, 00000028833HL2, 00000030875HL9, 00000030924HL5, 00000031916HL0, 00000031918HL6, 00000032909HL4, 00000033730HL3, 00000033974HL7, 00000034268HL3, 00000036165HL9, 00000042001HL8, 00000016177HL8, 00000031554HL9, 00000036626HL0, 00000013383HL5, 00000014323HL0, 00000019485HL2, 00000020232HL5, 00000010836HL5, 00000019679HL0, 00000023546HL5, 00000023547HL3, 00000030926HL0, 00000041996HL0, 00000042388HL9, 00000042537HL1, 00000011063HL5, 00000015391HL6, 00000017138HL9, 00000021058HL3, 00000030918HL7, 00000040086HL1, 00000010328HL3, 00000013375HL1, 00000014335HL4, 00000015065HL6, 00000015081HL3, 00000015747HL9, 00000016371HL7, 00000016701HL5, 00000017125HL6, 00000017481HL3, 00000019482HL9, 00000019486HL0, 00000019678HL2, 00000019680HL8, 00000019788HL9, 00000020081HL6, 00000020224HL2, 00000021639HL0, 00000022032HL7, 00000023118HL3, 00000023181HL1, 00000024432HL7, 00000029259HL9, 00000031908HL7, 00000031914HL5, 00000032231HL3, 00000032897HL1, 00000033724HL6, 00000033725HL3, 00000033968HL9, 00000034260HL0, 00000034459HL8, 00000034462HL2, 00000034783HL1, 00000035377HL1, 00000035378HL9, 00000036563HL5, 00000037254HL0, 00000038190HL5, 00000016175HL2, 00000019042HL1, 00000013796HL8, 00000011182HL3, 00000032313HL9, 00000012475HL0, 00000020222HL6, 00000029743HL2, 00000033726HL1, 00000020732HL4, 00000020737HL3, 00000012674HL8, 00000021551HL7, 00000031689HL3, 00000010651HL8, 00000015757HL8, 00000022520HL1, 00000015082HL1, 00000015153HL0, 00000021644HL0, 00000013373HL6, 00000015158HL9, 00000022606HL8, 00000022607HL6, 00000024489HL2, 00000024590HL2, 00000024671HL0, 00000024689HL2, 00000030446HL9, 00000030460HL0, 00000030927HL8, 00000036429HL9, 00000015387HL4, 00000039893HL3, 00000039895HL8, 00000039962HL6, 00000040002HL8, 00000040072HL1, 00000040164HL6, 00000040167HL9, 00000022402HL2, 00000036266HL5, 00000036914HL0, 00000037252HL4, 00000013564HL0.

Why it was recalled

The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.

Root cause (FDA determination)

Process control

Action the firm took

GE issued a Field Modification Instruction to their field service representatives, released in December 2007, with instructions to replace the labels on the console with the appropriate labeling. Consignees will be visited by GE field service representatives beginning December 2007. All consignees will have the required labels affixed to their units by July 31, 2008.

Recalling firm

Firm
GE Healthcare
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide Distribution, including USA, Canada, Mexico, France, Hong Kong, Italy, Korea, China, Poland, UK, New Zealand, Saudi Arabia, Taiwan, Turkey, and Australia.

Timeline

Recall initiated
2007-11-01
Posted by FDA
2008-06-10
Terminated
2011-12-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #68347. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as t · FDA Device Recalls