FDA Device Recalls

Recalls /

#69041

Product

Mevatron KD2 Medical Linear Accelerator, Material Number: 01940753, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens Medical Solutions USA, Inc, Concord, CA

FDA product code
IYEAccelerator, Linear, Medical
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K862339
Affected lot / code info
Serial Numbers: 2568 2698 2589 2038 2448 2547 2681 2686 2781 2592 and 2726

Why it was recalled

Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.

Root cause (FDA determination)

Software design

Action the firm took

This is a mandatory software upgrade for consignees. The firm's representatives have contacted consignees via Urgent Device Notification letters dated February 14, 2008, notifying them of the new update and making arrangements to install it.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
4040 Nelson Ave, Concord, California 94520-1200

Distribution

Distribution pattern
Worldwide-USA including states of AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MS, MO, MN, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI, and WY and countries of Angola, Australia, Austria,Belgium,Brazil,Canada,Chile,Colombia,Costa Rica, Croatia,Cyprus,The Czech Republic,Denmark,DPR of Korea,El Salvador,France,Germany,Greece,Hungary,India,Iran,Ireland,Italy,Japan,Jordan,Malaysia,Mexico, The Netherlands,New Zealand,Norway,P.R. China,Pakistan,Peru,Philippines,Poland,Portugal,Republic of Korea,Romania,Russian Fed.Saudi Arabia,Singapore,Slovakia,South Africa,Spain,Sweden,Taiwan,Thailand,Trinidad,Tobago,Turkey, U.A.E.,Ukraine, The United Kingdom, and Venezuela

Timeline

Recall initiated
2008-01-11
Posted by FDA
2008-04-02
Terminated
2011-05-06
Status

Source: openFDA Device Recall endpoint. Recall record ID #69041. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.