Product

Siemens Medical Solutions, Mevatron KD-2 Digital Linear Accelerator, part number 9822685 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV. Medical charged-particle radiation therapy system.

FDA product code
IYEAccelerator, Linear, Medical
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K862339
Affected lot / code info
Serial numbers:2348, 2224, 2431, 2131, and 2405

Why it was recalled

Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.

Root cause (FDA determination)

Device Design

Action the firm took

Siemens sent a Customer Safety Advisory Notice dated 12/12/2007 to DEVA customers. The letter stated that Siemens would temporarily disable energies 5MeV and below for all customers that decide to continue using the DEVA, resulting in the temporary unavailability of energies 5MeV and below for all available. Customers wishing not to have these energies blocked must return their DEVA and they would be contacted by a local Siemens rep to schedule a service appointment. The letter stated that as a preventative measure that those customers should refrain from using the DEVA with any combination of energies 5MeV and below with field sizes smaller than 6cm x 6cm. The firm suggested using the fixed field electron applicators and cutouts with these energies and field sizes. Siemens will release a new control console software in 2008 for free-of-charge, that will limit the use of the DEVA to energies 6MeV and above, while allowing the use of lower energies using the fixed field electron applicatiors.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
4040 Nelson Ave, Concord, California 94520-1200

Distribution

Distribution pattern
Worldwide Distribution: USA, Canada, Belgium, Bulgaria, France, Germany, Hungary, India, Italy, Japan, Malaysia, P.R. China, Pakistan, Poland, Portugal, Qatar, Republic Korea, South Africa, Spain, Sweden, and United Kingdom.

Timeline

Recall initiated
2007-12-10
Posted by FDA
2008-09-29
Terminated
2011-05-13
Status

Source: openFDA Device Recall endpoint. Recall record ID #69068. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.