Zimmer Spine Cyclone anterior cervical plate, 3 Level, 54MM (Ti-6Al-4V); REF 600-03-154.
- FDA product code
- KWQ — Appliance, Fixation, Spinal Intervertebral Body
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K031214, K040397
- Affected lot / code info
- All lots.
Why it was recalled
The screw locking cap may fracture when the surgeon rotates it into the locked position.
Root cause (FDA determination)
Device Design
Action the firm took
Consignees were notified via voicemail on 3/26/08 and a letter was sent to all distributors on 3/27/08 informing them that the plate should no longer be used and that all lots are being recalled. A letter dated 6/18/08 was issued to implant physicians recommending patient monitoring. Contact Zimmer Spine at 1-952-857-5636 for assistance.