Product
GE Healthcare Precision RXi version e Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, and 50KW or 65 kW pulsing generator. The system is configured with a 23 cm image intensifier and a 512x512 Image Chain.
- FDA product code
- JAA — System, X-Ray, Fluoroscopic, Image-Intensified
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K041605
- Affected lot / code info
- 10805 10805 10805 10804 10804 10804
Why it was recalled
GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .
Root cause (FDA determination)
Device Design
Action the firm took
Two letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect and correct the clearance between the console joystick and its mounting location. If your facility has a Precision RXi or Precision e system manufactured before March 2007, a GE Healthcare Field Engineer will schedule a field visit to install the added joystick protection. This upgrade will be performed at no charge. If you have any questions/concerns regarding this issue, please contact your local GE Healthcare sales or service representative.
Recalling firm
- Firm
- GE Healthcare
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- (USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE SI SK SY TN TR UA YU ZA
Timeline
- Recall initiated
- 2007-12-18
- Posted by FDA
- 2008-08-24
- Terminated
- 2012-08-09
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #70060. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.