FDA Device Recalls

Recalls /

#71765

Product

Biomet OSS Interlok IM Stem w/Screw, straight collar, 11 mm x 225 mm, TI6AL4V alloy, sterile, Biomet Orthopedics, Inc., Warsaw, IN; REF 150490. Orthopedic implant intended as part of a replacement knee joint.

FDA product code
KROProsthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K002757
Affected lot / code info
Lots 564030 and 217680.

Why it was recalled

Mislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while clarifying package contents.

Root cause (FDA determination)

Labeling False and Misleading

Action the firm took

On May 15, 2008 an Urgent Medical Device Removal Notice was issued to consignees and implanting physicians notifying them to return the product and of risks involved, respectively. Any adverse reactions which may be experienced with the use of this product should be reported to FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852, or the MedWatch website at www.fda.gov/medwatch. If you have any questions contact Mary Hardesty, Manager, Corrective Actions, Biomet Orthopedics, Inc. at 1-800-348-9500.

Recalling firm

Firm
Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582

Distribution

Distribution pattern
Class II Recall - Nationwide Distribution --- including states of California, Indiana, Kentucky, Massachusetts and Virginia.

Timeline

Recall initiated
2008-05-15
Posted by FDA
2008-09-11
Terminated
2008-11-03
Status