FDA Device Recalls

Recalls /

#71793

Product

Model 8709SC Indura 1P One-Piece Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.

FDA product code
LKKPump, Infusion, Implanted, Programmable
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P860004
Affected lot / code info
All lots numbers included

Why it was recalled

The SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.

Root cause (FDA determination)

Device Design

Action the firm took

Medtronic informed all Belgium implanting and managing physicians of the SynchroMed II infusion system. Each customer was mailed a Dear Healthcare Professional letter (Medtronic Urgent: Field Safety Notice) on 4/3/08 providing important safety information and patient management recommendations regarding the device incompatibility. The firm recommends that healthcare professionals do not use SC catheters if a connection with an implantable access port is required. Also, if a continuous infusion trial is required with use of an implantable access port, they should use a model 8711 intrathecal catheter. The firm states that for patients who are currently undergoing a continuous trial with an implantable port, replacing the catheter connector before connecting to the SynchroMedd II using the Medtronic model 8578 revision kit is recommended. Medtronic also recommends discussing the risks of connection failure with patients. Local field reps will personally follow-up with affected customers to confirm and document notification.

Recalling firm

Firm
Medtronic Neuromodulation
Address
800 53rd Ave NE PO Box 1250, Minneapolis, Minnesota 55440-1250

Distribution

Distribution pattern
International Distribution --- country of Belgium.

Timeline

Recall initiated
2008-04-14
Posted by FDA
2008-08-27
Terminated
2008-09-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #71793. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.