Product

Roche Diagnostics NAPA2, N-Acetyl-Procainamide, COBAS INTEGRA, Roche Diagnostics, Indianapolis, IN.; 20737852322. An in-vitro diagnostic reagent system intended for use on COBAS Integra systems for the qualitative determination of NAPA in serum or plasma.

FDA product code
CGXAlkaline Picrate, Colorimetry, Creatinine
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K951595
Affected lot / code info
Lot 60115501.

Why it was recalled

Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.

Root cause (FDA determination)

Labeling Change Control

Action the firm took

Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.

Recalling firm

Firm
Roche Diagnostics Corp.
Address
9115 Hague Rd, Indianapolis, Indiana 46256-1025

Distribution

Distribution pattern
Nationwide Distribution

Timeline

Recall initiated
2008-06-11
Posted by FDA
2008-10-09
Terminated
2009-09-22