FDA Device Recalls

Recalls /

#71908

Product

Access Myoglobin Reagents Kits, Part Number: 973243. The Access Myoglobin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Myoglobin levels in human serum and plasma using the Access Immunoassay Systems.

FDA product code
DDRMyoglobin, Antigen, Antiserum, Control
Device class
Class 2
Medical specialty
Immunology
510(k) numbers
K021229
Affected lot / code info
All Lot Numbers.

Why it was recalled

Beckman Coulter initiated the field action as a result of recent internal testing following the CLSI EP-5A - Evaluation of Precision Performance of Clinical Chemistry Devices protocol, it was determined that the total imprecision for the Access Myoglobin Reagent is greater than or equal to 10% across the expected physiological range rather than the stated claim less than 8%. Based on internal tes

Root cause (FDA determination)

Labeling False and Misleading

Action the firm took

An Urgent: Product Corrective Action (PCA) letter was mailed via US Postal Service on October 08, 2007 to the customers that received Access Myoglobin Reagent kits informing them about the issue. The letter also informed customers that the product insert in the Access Assay Manual will be revised with the new range to reflect the new imprecision claim. Customers were instructed to: (1) Share this information with their laboratory staff and retain the letter for their Quality System documentation, (2) Complete and return the enclosed fax-back response form within 10 days so that Beckman Coulter, Inc. can be assured that the customer received the notification, and (3) Contact Customer Technical Support at 1-800-854-3633 or your local Beckman Coulter, Inc. representative for questions or additional information regarding the notification.

Recalling firm

Firm
Beckman Coulter Inc
Address
200 S Kraemer Blvd, Brea, California 92822-6208

Distribution

Distribution pattern
Nationwide and Canada.

Timeline

Recall initiated
2007-09-27
Posted by FDA
2009-05-22
Terminated
2009-05-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #71908. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.