FDA Device Recalls

Recalls /

#72444

Product

COULTER LH SlideMaker, LH Barcode Specification, LH SM Intended Use, LH Slide and Sample Flow, Part Number 6605633, P/N 4277248, 4277299, and 277299. Intended for use as an optional peripheral to the Beckman Coulter LH 700 Series System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the LH 700 Series.

FDA product code
KPASlide Stainer, Automated
Device class
Class 1
Medical specialty
Pathology
Affected lot / code info
All Software Versions.

Why it was recalled

Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.

Root cause (FDA determination)

Pending

Action the firm took

An Urgent: Product Corrective Action letter, dated June 9, 2008, was sent to customers. The letter identified the affected products, stated the background, discussed the issue, and stated the actions to be taken. Customers are to: 1. Only use upper case alpha characters in the Sample Identifier 2. If you must use lower case alpha characters in the Sample Identifier, configure the SlideMaker Slide labels without a bar-code. Questions regarding the Product Corrective Action should be directed to Beckman Coulter Customer Service at 1-800-526-7694 or your local Beckman Coulter Rep.

Recalling firm

Firm
Beckman Coulter Inc
Address
200 S Kraemer Blvd, Brea, California 92822-6208

Distribution

Distribution pattern
Worldwide Distribution -- USA and Canada.

Timeline

Recall initiated
2008-05-29
Posted by FDA
2010-02-25
Terminated
2011-11-29
Status

Source: openFDA Device Recall endpoint. Recall record ID #72444. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.