Recalls / —
—#72457
Product
Zimmer Freestanding Patient Helper for Hill-Rom Totalcare beds; Zimmer Inc., Warsaw, IN; Prod. Nos. 00-2700-965-00 and 00-0965-001-00. The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed.
- FDA product code
- IKX — Aid, Transfer
- Device class
- Class 1
- Medical specialty
- Physical Medicine
- Affected lot / code info
- All units shipped between 7/11/07 and 5/19/08.
Why it was recalled
If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.
Root cause (FDA determination)
Device Design
Action the firm took
Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.
Recalling firm
- Firm
- Zimmer Inc.
- Address
- 345 E Main St, Warsaw, Indiana 46580-2746
Distribution
- Distribution pattern
- Nationwide and Canada.
Timeline
- Recall initiated
- 2008-07-07
- Posted by FDA
- 2008-09-16
- Terminated
- 2009-11-12
- Status
- —