Recalls / —
—#72782
Product
Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
- FDA product code
- KWS — Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K062250, K081620
- Affected lot / code info
- All lots.
Why it was recalled
Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.
Root cause (FDA determination)
Device Design
Action the firm took
A letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.
Recalling firm
- Firm
- Depuy Orthopaedics, Inc.
- Address
- 700 Orthopaedic Dr, Warsaw, Indiana 46582-3994
Distribution
- Distribution pattern
- Nationwide Distribution.
Timeline
- Recall initiated
- 2008-07-10
- Posted by FDA
- 2009-01-15
- Terminated
- 2009-10-28
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #72782. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.