FDA Device Recalls

Recalls /

#73323

Product

BICOR HS X-Ray System, Model number 6005776

FDA product code
IZISystem, X-Ray, Angiographic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K940484
Affected lot / code info
Model number 6005776 - Serial numbers: 3016 and 3026

Why it was recalled

Unintended movement of the system table and/or C arm.

Root cause (FDA determination)

Software design

Action the firm took

Siemens Medical Solutions issued a letter to consignees on 8/11/2008 to inform them of the problem and a workaround should the issue occur prior to software revision.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355

Distribution

Distribution pattern
Nationwide.

Timeline

Recall initiated
2008-08-11
Posted by FDA
2008-11-04
Terminated
2012-12-21
Status

Source: openFDA Device Recall endpoint. Recall record ID #73323. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.