FDA Device Recalls

Recalls /

#73600

Product

Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA and stent placement and atherectomies), pacemaker implantations and electrophysiology. It is compatible with specified magnetic navigation systems.

FDA product code
IZISystem, X-Ray, Angiographic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K041949
Affected lot / code info
Devices are identified with Site Numbers:  103281, 103283, 505903, 506119, 530653, 536715, 543411, 545396, 547823, 549368, 552969, 554802, 554970, 555293, 555299, 555834, 41443647, 41443804, 41444650, 41445448, 41445501, 41445538, 41455920, 41780160, 42057324, and 42064724.

Why it was recalled

Potential for table and stand not to move due to a software error.

Root cause (FDA determination)

Software design

Action the firm took

Philips Medical Systems North America Co. sent an URGENT - Device Correction, dated September 15, 2010, to their consignees. The letter identified the product, the problem, and the action to be taken. The consignees were informed that there was a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible. The consignees were advised that as geometry software errors happen intermittently, no actions can be performed by the consignees to minimize the risk except for not performing DRA procedures until the firm upgrades the software. For question consignees were to contact the Philips Call Center at 800-722-9377 #5, #4, #1 and reference FCO 72200111 or FCO 72200125.

Recalling firm

Firm
Philips Medical Systems North America Co. Phillips
Address
22100 Bothell Everett Hwy, Bothell, Washington 98021-8431

Distribution

Distribution pattern
Devices were distributed to hospitals and medical facilities through out the U.S.

Timeline

Recall initiated
2008-08-21
Posted by FDA
2010-10-22
Terminated
2010-11-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #73600. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.