FDA Device Recalls

Recalls /

#73623

Product

Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2.

FDA product code
LLZSystem, Image Processing, Radiological
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K052461, K071114
Affected lot / code info
Model number 10014063. Serial numbers: 1005, 1039, 1040, 1041, 1049, 1053, 1055, 1056, 1057, 1058, 1061, 1064, 1065, 1067, 1071, 1074, 1076, 1079, 1080, 1081, 1082, 1083, 1084, 1086, 1089, 1090, 1095, 1105, 1107, 1112, 1116, 1119, 1124, 1174, 1188, 1198, 1202, 1203, 1210, 1218, 1226, 1227, 1229, 1240, 1246, 1247, 1258, 1269, 1271, 1278, 1279, 1282, 1316, 1322, 1329, 1334, 1343, 1351, and 1353.

Why it was recalled

Image may not be visible after merge.

Root cause (FDA determination)

Software design

Action the firm took

The recalling firm issued Customer Safety Advisory Notice dated 8/22/08 to inform the consignees of the potential issue and provide instruction to avoid its occurrence. To avoid potential risk of this issue, users are asked to check whether the extended HL7 messages as stated in the letter are mandatory for their RIS-PACS workflow, and disable these messages if they are not. Siemens is preparing a modification under the UI IM086/08/S for the affected versions of syngo Imaging that will resolve the potential malfunction. The field modification will be immediately distributed and installed by the firm's Service as soon as it is available.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355

Distribution

Distribution pattern
Nationwide Distribution --- including states of AL, AZ, CA, CO, CT, FL, IA, KS, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, TN, TX, and WI.

Timeline

Recall initiated
2008-08-22
Posted by FDA
2008-10-15
Terminated
2008-11-19
Status