FDA Device Recalls

Recalls /

#73830

Product

Triathlon Total Knee system Primary Tibial Baseplate #5 Beaded wPA; Catalog Number 5526-B-500, Cementless use only. Sterile. Howmedica Osteonics, 325 Corporate Drive, Mahwah, NJ 07430; A subsidiary of Stryker Corp. Made in Ireland. The Triathlon Total Knee System is intended for use in primary and revision total knee arthroplasty to alleviate pain and restore function.

FDA product code
MBHProsthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K051380
Affected lot / code info
Lot Code SANMH

Why it was recalled

Stryker Orthopaedics became aware that one lot of the Triathlon Primary Beaded PA Baseplate Size 5 may not have the Peri-Apatite coating.

Root cause (FDA determination)

Process change control

Action the firm took

Important Market Withdrawal notification letters were sent via Fed Ex on 9/20/06 to the direct accounts. Notification letters informed of the reason for the market withdrawal and requested that all branches examine their inventory and hospital consignment locations to identify the product. Product was to be reconciled on the Product Accountability Form and returned to Stryker as per instructions. Contact Stryker Howmedica Osteonics Corp at 1-201-831-5718 for assistance.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Dr, Mahwah, New Jersey 07430

Distribution

Distribution pattern
PA, GA, AL, WI, UT, and OR.

Timeline

Recall initiated
2006-09-21
Posted by FDA
2009-01-08
Terminated
2009-01-09
Status