Recalls / —
—#73830
Product
Triathlon Total Knee system Primary Tibial Baseplate #5 Beaded wPA; Catalog Number 5526-B-500, Cementless use only. Sterile. Howmedica Osteonics, 325 Corporate Drive, Mahwah, NJ 07430; A subsidiary of Stryker Corp. Made in Ireland. The Triathlon Total Knee System is intended for use in primary and revision total knee arthroplasty to alleviate pain and restore function.
- FDA product code
- MBH — Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K051380
- Affected lot / code info
- Lot Code SANMH
Why it was recalled
Stryker Orthopaedics became aware that one lot of the Triathlon Primary Beaded PA Baseplate Size 5 may not have the Peri-Apatite coating.
Root cause (FDA determination)
Process change control
Action the firm took
Important Market Withdrawal notification letters were sent via Fed Ex on 9/20/06 to the direct accounts. Notification letters informed of the reason for the market withdrawal and requested that all branches examine their inventory and hospital consignment locations to identify the product. Product was to be reconciled on the Product Accountability Form and returned to Stryker as per instructions. Contact Stryker Howmedica Osteonics Corp at 1-201-831-5718 for assistance.
Recalling firm
- Firm
- Stryker Howmedica Osteonics Corp.
- Address
- 325 Corporate Dr, Mahwah, New Jersey 07430
Distribution
- Distribution pattern
- PA, GA, AL, WI, UT, and OR.
Timeline
- Recall initiated
- 2006-09-21
- Posted by FDA
- 2009-01-08
- Terminated
- 2009-01-09
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #73830. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.