Product

Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular P Module, Roche Diagnostics Corp., Indpls., IN; GMMI No. 03284549001. Is a fully automatic, computerized chemistry analyzer intended for in vitro quantitative and qualitative determinations of a wide range analytes. The instrument is capable of performing potentiometric and photometric assays.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K953239
Affected lot / code info
All units with software version 08-03.

Why it was recalled

If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.

Root cause (FDA determination)

Software change control

Action the firm took

Consignees were notified via an Urgent Medical Device Correction letter dated 11/12/08 and instructed to only execute "module unmasking" when the unit is in "standby" mode and to consult with their physicians at the facility to determine the clinical implications and whether any clinical intervention is needed for patients treated based upon results obtained with this software. Questions should be directed to Roche Diagnostics Technical Support at 1-800-428-2336.

Recalling firm

Firm
Roche Diagnostics Corp.
Address
9115 Hague Rd, Indianapolis, Indiana 46256-1025

Distribution

Distribution pattern
Nationwide Distribution -- including states of California, Florida, Georgia, Hawaii, Indiana, Maryland, New York, North Carolina, Pennsylvania, and Virginia.

Timeline

Recall initiated
2008-11-12
Posted by FDA
2009-04-10
Terminated
2009-09-22
Status