Recalls / —
—#75698
Product
Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico
- FDA product code
- HRX — Arthroscope
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K944210
- Affected lot / code info
- Lot numbers 08135AG2 through 08294AG2.
Why it was recalled
Leak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed.
Root cause (FDA determination)
Component change control
Action the firm took
Recall initiated 12/01/2008, Sales representatives were notified via telephone and e-mail. Product advisories entitled ""Urgent: Device Removal" and dated December 1, 2008, were sent to all consignees. All unused product is to be returned to the firm. For further information, please contact Stryker Endoscopy by telephone at 408-754-2124.
Recalling firm
- Firm
- Stryker Endoscopy
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide Distribution --- including USA and countries of Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, New Zealand, Poland, Portugal, Sweden, Spain, Singapore, Switzerland, and UK.
Timeline
- Recall initiated
- 2008-12-01
- Posted by FDA
- 2009-01-16
- Terminated
- 2011-01-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #75698. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.