Recalls / —
—#78879
Product
Innova 4100 / 4100 IQ, GE Healthcare, Waukesha, WI 53188 ( The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). Indicated for use in diagnostic and interventional angiographic procedures of human anatomy. Intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. Not intended for mammography applications.
- FDA product code
- MQB — Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K023178, K042053
- Affected lot / code info
- 757398INNOVA
Why it was recalled
GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to recover.
Root cause (FDA determination)
Software design
Action the firm took
On site visits were performed from 3/6/08 - 12/19/08 to update software on all 5 units.
Recalling firm
- Firm
- GE Medical Systems, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide Distribution -- USA including states of MD, TX, and VA and country of China.
Timeline
- Recall initiated
- 2008-03-06
- Posted by FDA
- 2009-03-12
- Terminated
- 2014-09-04
- Status
- —