Recalls / —
—#79482
Product
AxSYM System Tubing Decontamination Solution; 500 mL bottle; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064; List Number 07B05-09. To be used only for cleaning and inhibition of fouling bacteria in the tubing system of the in vitro immunoassay analyzer. This product does not disinfect or sterilize the tubing system.
- FDA product code
- JJE — Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K974651
- Affected lot / code info
- List 7B05-09, Lot Number: 211974, expiration date 9/30/09.
Why it was recalled
Bottle labels from lot 211974, AxSYM System Tubing Decontamination Solution (List Number 7B05-09), were found to be missing the following information: lot number, date of manufacture, and expiration date.
Root cause (FDA determination)
Manufacturing material removal
Action the firm took
Abbott sent Product Recall Immediate Action Required letters dated February 18, 2009 to customers informing them of the affected product. Customers were instructed to examine their inventory and dispose all affected product. Accounts were requested to complete and fax back customer reply sheet acknowledging receipt of the letter, as well as the number of bottles destroyed, and to retain a copy of the letter for their records. For further information, contact Abbott Customer Service at 1-877-4ABBOTT.
Recalling firm
- Firm
- Abbott Laboratories
- Address
- 100 Abbott Park Rd, Abbott Park, Illinois 60064-3502
Distribution
- Distribution pattern
- Worldwide Distribution -- United States including Puerto Rico and Argentina, Brazil, Canada, Chile, Columbia, Germany, Hong Kong, Japan, Mexico, Peru, South Korea, Thailand and Trinidad & Tobago.
Timeline
- Recall initiated
- 2009-02-18
- Posted by FDA
- 2009-09-30
- Terminated
- 2010-07-08
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #79482. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.