FDA Device Recalls

Recalls /

#79482

Product

AxSYM System Tubing Decontamination Solution; 500 mL bottle; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064; List Number 07B05-09. To be used only for cleaning and inhibition of fouling bacteria in the tubing system of the in vitro immunoassay analyzer. This product does not disinfect or sterilize the tubing system.

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K974651
Affected lot / code info
List 7B05-09, Lot Number: 211974, expiration date 9/30/09.

Why it was recalled

Bottle labels from lot 211974, AxSYM System Tubing Decontamination Solution (List Number 7B05-09), were found to be missing the following information: lot number, date of manufacture, and expiration date.

Root cause (FDA determination)

Manufacturing material removal

Action the firm took

Abbott sent Product Recall Immediate Action Required letters dated February 18, 2009 to customers informing them of the affected product. Customers were instructed to examine their inventory and dispose all affected product. Accounts were requested to complete and fax back customer reply sheet acknowledging receipt of the letter, as well as the number of bottles destroyed, and to retain a copy of the letter for their records. For further information, contact Abbott Customer Service at 1-877-4ABBOTT.

Recalling firm

Firm
Abbott Laboratories
Address
100 Abbott Park Rd, Abbott Park, Illinois 60064-3502

Distribution

Distribution pattern
Worldwide Distribution -- United States including Puerto Rico and Argentina, Brazil, Canada, Chile, Columbia, Germany, Hong Kong, Japan, Mexico, Peru, South Korea, Thailand and Trinidad & Tobago.

Timeline

Recall initiated
2009-02-18
Posted by FDA
2009-09-30
Terminated
2010-07-08
Status