FDA Device Recalls

Recalls /

#79684

Product

Pilling(R) Coronary Scissors, 60Deg 7", Catalog numbers: 352166, Teleflex Medical, Research Triangle Park, NC 27709

FDA product code
LRWScissors, General, Surgical
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Date Codes: RR8, SS8, TT8, UU8, VV8, WW8, XX8, and MM9.

Why it was recalled

The product has a rough square finish instead of a smooth rounded finish which can cause damage to arteries.

Root cause (FDA determination)

Process control

Action the firm took

Consignees were notified by an Urgent Medical Device Recall letter, dated 03/03/2009, via UPS Next Day. Hospitals were instructed to immediately discontinue use of the products, inspect the scissors for the date codes and to return the product to Teleflex Medical for replacement. Distributors were instructed to forward the letter to their consignees to retrieve relevant product from those locations. A Recall Acknowledgement Form was attached to be completed and faxed back to Teleflex Medical. A second mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or they may be contacted by phone, e-mail or fax. A record of these contacts will be maintained. Affected product retrieved by Teleflex Medical will be destroyed and replacement product provided to the consignee. Records of destruction will be maintained.

Recalling firm

Firm
Teleflex Medical
Address
4024 Stirrup Creek Dr, Durham, North Carolina 27703-9000

Distribution

Distribution pattern
Worldwide Distribution -- USA including states of AL, AZ, CA, CO, DC, DE, FL, IA, IN, KS, MA, MD, MI, MO, MS, NH, NJ, NY, NV, OH, OK, OR, PA, RI, TN, TX, UT, and WA, and countries of Australia, Canada, France, Germany, India, Japan, and Singapore.

Timeline

Recall initiated
2009-03-03
Posted by FDA
2009-04-09
Terminated
2011-03-31
Status

Source: openFDA Device Recall endpoint. Recall record ID #79684. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.