FDA Device Recalls

Recalls /

#80290

Product

Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-5, 2377062-30)

FDA product code
LNHSystem, Nuclear Magnetic Resonance Imaging
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K032795
Affected lot / code info
Model Number: 2138300-30;  Serial Numbers: 00000203387MR6, 00000208208MR9, 00000216409MR3, 00000229302MR5, 00000229306MR6, 00000230309MR7, 00000231588MR5, 00000237930MR3, 00000239847MR7, 00000241694MR9, 00000247255MR3, 00000253103MR6, 00000189633MR1, 00000192104MR8, 00000193652MR5, 00000194218MR4, 00000194219MR2, 00000194950MR2, 00000196364MR4, 00000197334MR6, 00000198709MR8, 00000199443MR3, 00000201708MR5, 00000201709MR3, 00000201710MR1, 00000201901MR6, 00000202223MR4, 00000203386MR8, 00000204837MR9, 00000204838MR7, 00000204839MR5, 00000206438MR4, 00000206439MR2, 00000207493MR8, 00000207494MR6, 00000208209MR7, 00000208210MR5, 00000209712MR9, 00000209713MR7, 00000212511MR0, 00000212513MR6, 00000212514MR4, 00000212515MR1, 00000215366MR6, 00000216405MR1, 00000216406MR9, 00000216407MR7, 00000216409MR3, 00000217198MR1, 00000217201MR3, 00000221250MR4, 00000221251MR2, 00000221253MR8, 00000221254MR6, 00000222950MR8, 00000225273MR2, 00000225274MR0, 00000225942MR2, 00000226318MR4, 00000228022MR0, 00000228023MR8, 00000228024MR6, 00000228773MR8, 00000229303MR3, 00000229304MR1, 00000229305MR8, 00000230308MR9, 00000230312MR1, 00000231591MR9, 00000231592MR7, 00000232328MR5, 00000233644MR4, 00000233646MR9, 00000234832MR4, 00000234833MR2, 00000234834MR0, 00000237051MR8, 00000237052MR6, 00000237053MR4, 00000237926MR1, 00000237927MR9, 00000237928MR7, 00000237929MR5, 00000239480MR7, 00000239482MR3, 00000239483MR1, 00000239484MR9, 00000239845MR1, 00000239846MR9, 00000241126MR2, 00000241127MR0, 00000241693MR1, 00000241695MR6, 00000243405MR8, 00000244231MR7, 00000244232MR5, 00000247252MR0, 00000247253MR8, 00000247254MR6, 00000247256MR1, 00000250184MR9, 00000250185MR6, 00000250186MR4, 00000250187MR2, 00000253039MR2, 00000253102MR8, Unknown;  Model Number: 2377062-5;  Serial Numbers: 00000000100YR8, 00000000119YR8, 00000000129YR7, 00000000102YR4, 00000000103YR2, 00000000105YR7, 00000000106YR5, 00000000108YR1, 00000000109YR9, 00000000111YR5, 00000000113YR1, 00000000118YR0, 00000000120YR6, 00000000122YR2, 00000000124YR8, 00000000127YR1, 00000000128YR9, 00000000131YR3, 00000000132YR1, 00000000133YR9, 00000000134YR7, 00000000135YR4, 00000000136YR2, 00000000137YR0, 00000000142YR0, 00000000144YR6, 00000000146YR1, 00000000149YR5, 00000000150YR3, 00000000151YR1, 00000000152YR9, 00000000155YR2, 00000943610YM6, 00000951021YM5, 00000962592YM2;  Model Number: 2377062-30  Serial Numbers: 00000003047YR8, 00000003011YR4, 00000003022YR1, 00000003023YR9, 00000003024YR7, 00000003025YR4, 00000003026YR2, 00000003027YR0, 00000003028YR8, 00000003029YR6, 00000003030YR4, 00000003031YR2, 00000003033YR8, 00000003034YR6, 00000003035YR3, 00000003036YR1, 00000003039YR5, 00000003041YR1, 00000003042YR9, 00000003043YR7 and 00000003044YR5.

Why it was recalled

GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. Two injuries and four complaints filed. 2/20/09: Recall being expanded to incl

Root cause (FDA determination)

Device Design

Action the firm took

GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Healthcare sales or service representative.

Recalling firm

Firm
GE Medical Systems, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL REPUBLIC OF FRANCE, GET 7 D CHINA, GREECE, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, MALAWI, MEXICO, OMAN, PANAMA, PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SPAIN , SWITZERLAND, TURKEY, UKRAINE, UNITED ARAB EMERITUS, UNITED KINGDOM and VENEZUELA.

Timeline

Recall initiated
2008-11-03
Posted by FDA
2009-08-25
Terminated
2012-05-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #80290. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.