Recalls / —
—#80290
Product
Signa Openspeed 0.7T MR Systems, Model Numbers: Openspeed 3 (2138300-30), Openspeed 4 (2377062-5), Openspeed 5 (2377062-5, 2377062-30)
- FDA product code
- LNH — System, Nuclear Magnetic Resonance Imaging
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K032795
- Affected lot / code info
- Model Number: 2138300-30; Serial Numbers: 00000203387MR6, 00000208208MR9, 00000216409MR3, 00000229302MR5, 00000229306MR6, 00000230309MR7, 00000231588MR5, 00000237930MR3, 00000239847MR7, 00000241694MR9, 00000247255MR3, 00000253103MR6, 00000189633MR1, 00000192104MR8, 00000193652MR5, 00000194218MR4, 00000194219MR2, 00000194950MR2, 00000196364MR4, 00000197334MR6, 00000198709MR8, 00000199443MR3, 00000201708MR5, 00000201709MR3, 00000201710MR1, 00000201901MR6, 00000202223MR4, 00000203386MR8, 00000204837MR9, 00000204838MR7, 00000204839MR5, 00000206438MR4, 00000206439MR2, 00000207493MR8, 00000207494MR6, 00000208209MR7, 00000208210MR5, 00000209712MR9, 00000209713MR7, 00000212511MR0, 00000212513MR6, 00000212514MR4, 00000212515MR1, 00000215366MR6, 00000216405MR1, 00000216406MR9, 00000216407MR7, 00000216409MR3, 00000217198MR1, 00000217201MR3, 00000221250MR4, 00000221251MR2, 00000221253MR8, 00000221254MR6, 00000222950MR8, 00000225273MR2, 00000225274MR0, 00000225942MR2, 00000226318MR4, 00000228022MR0, 00000228023MR8, 00000228024MR6, 00000228773MR8, 00000229303MR3, 00000229304MR1, 00000229305MR8, 00000230308MR9, 00000230312MR1, 00000231591MR9, 00000231592MR7, 00000232328MR5, 00000233644MR4, 00000233646MR9, 00000234832MR4, 00000234833MR2, 00000234834MR0, 00000237051MR8, 00000237052MR6, 00000237053MR4, 00000237926MR1, 00000237927MR9, 00000237928MR7, 00000237929MR5, 00000239480MR7, 00000239482MR3, 00000239483MR1, 00000239484MR9, 00000239845MR1, 00000239846MR9, 00000241126MR2, 00000241127MR0, 00000241693MR1, 00000241695MR6, 00000243405MR8, 00000244231MR7, 00000244232MR5, 00000247252MR0, 00000247253MR8, 00000247254MR6, 00000247256MR1, 00000250184MR9, 00000250185MR6, 00000250186MR4, 00000250187MR2, 00000253039MR2, 00000253102MR8, Unknown; Model Number: 2377062-5; Serial Numbers: 00000000100YR8, 00000000119YR8, 00000000129YR7, 00000000102YR4, 00000000103YR2, 00000000105YR7, 00000000106YR5, 00000000108YR1, 00000000109YR9, 00000000111YR5, 00000000113YR1, 00000000118YR0, 00000000120YR6, 00000000122YR2, 00000000124YR8, 00000000127YR1, 00000000128YR9, 00000000131YR3, 00000000132YR1, 00000000133YR9, 00000000134YR7, 00000000135YR4, 00000000136YR2, 00000000137YR0, 00000000142YR0, 00000000144YR6, 00000000146YR1, 00000000149YR5, 00000000150YR3, 00000000151YR1, 00000000152YR9, 00000000155YR2, 00000943610YM6, 00000951021YM5, 00000962592YM2; Model Number: 2377062-30 Serial Numbers: 00000003047YR8, 00000003011YR4, 00000003022YR1, 00000003023YR9, 00000003024YR7, 00000003025YR4, 00000003026YR2, 00000003027YR0, 00000003028YR8, 00000003029YR6, 00000003030YR4, 00000003031YR2, 00000003033YR8, 00000003034YR6, 00000003035YR3, 00000003036YR1, 00000003039YR5, 00000003041YR1, 00000003042YR9, 00000003043YR7 and 00000003044YR5.
Why it was recalled
GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. Two injuries and four complaints filed. 2/20/09: Recall being expanded to incl
Root cause (FDA determination)
Device Design
Action the firm took
GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Healthcare sales or service representative.
Recalling firm
- Firm
- GE Medical Systems, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL REPUBLIC OF FRANCE, GET 7 D CHINA, GREECE, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, MALAWI, MEXICO, OMAN, PANAMA, PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SPAIN , SWITZERLAND, TURKEY, UKRAINE, UNITED ARAB EMERITUS, UNITED KINGDOM and VENEZUELA.
Timeline
- Recall initiated
- 2008-11-03
- Posted by FDA
- 2009-08-25
- Terminated
- 2012-05-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #80290. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.