Recalls / —
—#80863
Product
AXIOM Sireskop SD Model numbers 8890415 and 8890407. The product is intended for use as radiographic and fluoroscopic imaging device.
- FDA product code
- MQB — Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K051602
- Affected lot / code info
- Serial numbers: 51010, 71136, 51007, 51008, 71019, 71080, 71172, 71067, 41012, 71015, 71038, 71102, 71104, 71169, 51006, 51011, 51024, 51028, 71026, 71039, 71083, 71117, 71119, 41027, 71120, 71121, 41028, 71137, 71138, 71149, 71158, 71161, 71167, 71168, 71170, 51060, 71160, 51026, 71030, 71059, 71033, 71037, 41006, 51052, 51064, 71001, 71082, 71114, 71141, 71148, 71152, 71156, 51046, 71181, 51013, 51044, 71075, 71098, 51054, 71020, 71021, 71074, 71088, 41026, 71135, 71140, 71055, 71006, 71017, 71042, 71073, 71162, 71174, 41003, 51049, 71053, 71128, 71131, 41022, 71096, 71099, 71129, 51009, 51036, 51045, 71069, 71087, 71144, 41008, 71014, 71016, 71028, 71109, 71130, 71004, 71051, 71090, 71157, 51029, 51030, 51069, 71009, 71011, 71034, 71035, 71126, 71127, 71145, 51003, 51016, 71022, 71029, 71164, 51043, 51070, 71061, 71064, 71110, 71116, 51001, 51057, 41014, 71124, 71133, 51012, 51050, 71070, 71142, 71095, 41011, 51047, 71023, 71076, 71081, 71084, 51041, 41010, 71032, 41029, 71002, 71065, 71105, 71115, 51005, 71068, 51017, 51018, 51023, 71043, 71046, 41020, 71091, 71092, 71101, 71108, 71139, 71146, 71150, 71155, 71178, 51015, 51059, 51066, 71013, 71057, 71062, 71097, 41013, 71024, 41023, 71086, 51037, 71010, 71106, 71112, 51065, 51068, 71008, 41017, 71078, 41024, 71079, 71111, 71018, 71147, 51002, 71040, 41002, 51019, 51020, 41005, 51033, 41007, 51040, 51048, 51051, 51055, 51061, 71003, 71007, 71012, 41018, 71025, 41019, 71045, 71063, 71100, 71103, 71113, 71122, 71125, 71153, 71154, 71159, 71163, 71171, 71173, 71177, 51004, 51021, 51025, 51058, 71048. 71058, 71060, 71072, 41001, 51022, 51031, 51035, 51039, 71027, 71175, 51063, 71005, 71066, 71089, 71165, 71166, 71123, 71151, and 51056.
Why it was recalled
During interventional use and applications, the possibility exists for liquids to enter the systems.
Root cause (FDA determination)
Equipment maintenance
Action the firm took
A Customer Safety Advisory Notice dated January 8, 2009 was issued to affected customers via Update Instructions AX069/08/S. This letter informed customers of the potential issues and provided instructions to avoid its occurrence. The firm has released an Addendum to the User Manual warning of the potential hazards if liquids penetrate the systems during interventional use and applications. Direct questions about this recall to Siemens Medical Solutions USA, Inc. by calling 1-610-219-6300.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 51 Valley Stream Pkwy, Malvern, Pennsylvania 19355
Distribution
- Distribution pattern
- Nationwide Distribution.
Timeline
- Recall initiated
- 2009-01-08
- Posted by FDA
- 2009-05-06
- Terminated
- 2009-09-01
- Status
- —