FDA Device Recalls

Recalls /

#81661

Product

TransFx External Fixation System Drill Bit, Quick Connect, 4.0 mm diameter, 230 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-67. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

FDA product code
LXTAppliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K003205, K984357, K990848, K991723
Affected lot / code info
Lots 53012100, 53012200, 53022800, 53023100, 56023500, 70186300, 70239200 and 70247800.

Why it was recalled

Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

Root cause (FDA determination)

Process control

Action the firm took

Distributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.

Recalling firm

Firm
Zimmer Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
Worldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.

Timeline

Recall initiated
2009-05-01
Posted by FDA
2009-06-30
Terminated
2010-11-03
Status