Recalls / —
—#81877
Product
Zimmer NexGen Complete Knee Solution Legacy Knee- Posterior Stabilized LPS Femoral Component Option Size D Left, sterile, Zimmer, Warsaw, IN; REF 00-5996-014-51.
- FDA product code
- NJL — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P060037
- Affected lot / code info
- Lot 61207901.
Why it was recalled
The implant surface may not have been polished adequately, which could result in increased wear and polyethylene debris generation.
Root cause (FDA determination)
Employee error
Action the firm took
Consignees were notified by phone beginning on 4/17/09 and email and letter notifications were sent on 4/22/09. A dear surgeon letter entitled Urgent: Medical Device Recall, dated April 28, 2009, is being sent to implanting surgeons recommending patient monitoring. Consignees are to return the response form indicating their receipt of the notification. Questions or concerns should be directed to (800) 846-4637 or www.zimmer.com.
Recalling firm
- Firm
- Zimmer Inc.
- Address
- 345 E Main St, Warsaw, Indiana 46580-2746
Distribution
- Distribution pattern
- Worldwide Distribution -- USA including states of Arizona, California, Colorado, Hawaii, Illinois, Indiana, Michigan, Minnesota, Mississippi, New Jersey, New Mexico, New York, Ohio, Oregon, and Pennsylvania and country of Canada.
Timeline
- Recall initiated
- 2009-04-17
- Posted by FDA
- 2009-06-25
- Terminated
- 2009-11-12
- Status
- —