FDA Device Recalls

Recalls /

#82007

Product

GE Healthcare, GE Innova 2100 IQ Versatile Cardiac and Vascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

FDA product code
IZISystem, X-Ray, Angiographic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K050489, K052157, K052412, K061163
Affected lot / code info
SYSTEM ID:  850060731 853060205 82416020006 82416090003 82416120015 82416120016 82416210014 83016022101218 83016842101519 0002780363 00052VAS04  00052VAS05  00060VAS03  010120RX44 082416030012 082416130029 083016002001498 083016802101749 083016822101709 083016862101268 0910064010 110019RX13  1299563 17005VAS01  304526INNOVA  318449INNOVA3  407321INCL2  410368INNOVA1  503561CL1  503561CL4  526354XR13 573815BCR3  603882CATH1  609757IN2100  706774HT4 717217INN21 717544GECATH2  732776CATH4 732923INN 816347SLE2100 816861KCVAI2 816861KCVAI2  818INNOVAHMS1  828315FEP  910343NHIN3  A5152993 A5330714 B5483438 BG4033VA01 C5123626  GOLD010016 HR4106VA01 HR4107VA01 HU1173VA03 LV4000VA04 M4183841 M4484974 RO1180VA01 RU3301VA01 YV0562

Why it was recalled

Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra

Root cause (FDA determination)

Device Design

Action the firm took

Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.

Recalling firm

Firm
GE Medical Systems, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.

Timeline

Recall initiated
2009-04-16
Posted by FDA
2009-11-09
Terminated
2015-12-03
Status

Source: openFDA Device Recall endpoint. Recall record ID #82007. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.