Product
Product's marketing brochure is labeled in part: "Philips Easy Diagnost Eleva Conventional R/F system...PHILIPS". manufactured by Philips Medical Systems Gmbh, Hamburg, Germany.
- FDA product code
- KPR — System, X-Ray, Stationary
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K031535
- Affected lot / code info
- Site Numbers: 506250, 506251, 506276, 519152, 520209, 521153, 530573, 530607, 532426, 532779, 533491, 533574, 534207, 534419, 534644, 534647, 534675, 534803, 534945, 535378, 535437, 535785, 536051, 536458, 536522, 536636, 536637, 536687, 536765, 536840, 536933, 536941, 536973, 536985, 536994, 537013, 537087, 537091, 537179, 537186, 537492, 537518, 537571, 537689, 537761, 537796, 537803, 537863, 537945, 537946, 537970, 537971, 538237, 538328, 538351, 538407, 538451, 538458, 538462, 538477, 538480, 538519, 538759, 538812, 538814, 538866, 538891, 538901, 539040, 539095, 539101, 539103, 539237, 539291, 539310, 539339, 539377, 539378, 539379, 539411, 539419, 539488, 539501, 539519, 539527, 539529, 539530, 539557, 539590, 539683, 539785, 540022, 540163, 540187, 540361, 540380, 540445, 540457, 540459, 540624, 540788, 540810, 541011, 541054, 541063, 541078, 541083, 541093, 541348, 541402, 541494, 541535, 541614, 541694, 541721, 541803, 541838, 541850, 541872, 541884, 542061, 542121, 542122, 542205, 542308, 542401, 542521, 542537, 542589, 542593, 542716, 542829, 542849, 543136, 543144, 543224, 543241, 543249, 543275, 543297, 543331, 543369, 543415, 543426, 543619, 543620, 543660, 543723, 543754, 543755, 543771, 543776, 543803, 543841, 544075, 544113, 544172, 544228, 544236, 544297, 544341, 544417, 544480, 544481, 544568, 544648, 544654, 544695, 544937, 544999, 545000, 545038, 545120, 545163, 545166, 545213, 545231, 545234, 545244, 545391, 545452, 545710, 545746, 545780, 545783, 545824, 545844, 545923, 545948, 545951, 546029, 546155, 546490, 546524, 546551, 546583, 546602, 547054, 547189, 547243, 547297, 547304, 547371, 547598, 547599, 547655, 547657, 547744, 547758, 547759, 547798, 547936, 547937, 548021, 548276, 548290, 548302, 548382, 548559, 548784, 549008, 549080, 549269, 549270, 549381, 549416, 549509, 549557, 549616, 549663, 550072, 550079, 550197, 550313, 550591, 550804, 550936, 550960, 551005, 551416, 551573, 551616, 551645, 551721, 551759, 551941, 551954, 551980, 552204, 552205, 552211, 552324, 552363, 552364, 552407, 552502, 552579, 552594, 552626, 552640, 552917, 552918, 552965, 552966, 553013, 553172, 553207, 553224, 553230, 553275, 553349, 553387, 553441, 553448, 553455, 553563, 553850, 553893, 553900, 553926, 554061, 554483, 554581, 554641, 554721, 554722, 554769, 555052, 555269, 555458, 555524, 555574, 555626, 555964, 555993, 556124, 556621, 556622, 556726, 556785, 556984, 557177, 557259, 557272, 558437, 558596, 558734, 558775, 41443799, 41444441, 41658234, 41862667, 42039161, and 537967 / 49371595.
Why it was recalled
Patient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked to the incorrect patient, which may adversely affect diagnosis and treatment.
Root cause (FDA determination)
Software design
Action the firm took
Philips Medical Systems issued URGENT-Device Correction letters dated January 27, 2009 to consignees describing the issue with the Easy Diagnost Eleva GXR RF systems. The consignees were informed of the potential for incorrect diagnosis or treatment if an incorrect image is used for diagnosis. Consignees should contact the Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FC0 70600030" with any communication.
Recalling firm
- Firm
- Philips Medical Systems North America Co. Phillips
- Address
- 22100 Bothell Everett Hwy, Bothell, Washington 98021-8431
Distribution
- Distribution pattern
- Nationside Distribution: USA, including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PA, RI, SC, SD, TN, TX, TX, UT, VA, VT, WA, WI, and WV. Devices are not distributed to any foreign consignees.
Timeline
- Recall initiated
- 2009-01-21
- Posted by FDA
- 2010-11-26
- Terminated
- 2010-12-30
- Status
- —