FDA Device Recalls

Recalls /

#82093

Product

Product's marketing brochure is labeled in part: "Philips Easy Diagnost Eleva Conventional R/F system...PHILIPS". manufactured by Philips Medical Systems Gmbh, Hamburg, Germany.

FDA product code
KPRSystem, X-Ray, Stationary
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K031535
Affected lot / code info
Site Numbers:  506250, 506251, 506276, 519152, 520209, 521153, 530573, 530607, 532426, 532779, 533491, 533574, 534207, 534419, 534644, 534647, 534675, 534803, 534945, 535378, 535437, 535785, 536051, 536458, 536522, 536636, 536637, 536687, 536765, 536840, 536933, 536941, 536973, 536985, 536994, 537013, 537087, 537091, 537179, 537186, 537492, 537518, 537571, 537689, 537761, 537796, 537803, 537863, 537945, 537946, 537970, 537971, 538237, 538328, 538351, 538407, 538451, 538458, 538462, 538477, 538480, 538519, 538759, 538812, 538814, 538866, 538891, 538901, 539040, 539095, 539101, 539103, 539237, 539291, 539310, 539339, 539377, 539378, 539379, 539411, 539419, 539488, 539501, 539519, 539527, 539529, 539530, 539557, 539590, 539683, 539785, 540022, 540163, 540187, 540361, 540380, 540445, 540457, 540459, 540624, 540788, 540810, 541011, 541054, 541063, 541078, 541083, 541093, 541348, 541402, 541494, 541535, 541614, 541694, 541721, 541803, 541838, 541850, 541872, 541884, 542061, 542121, 542122, 542205, 542308, 542401, 542521, 542537, 542589, 542593, 542716, 542829, 542849, 543136, 543144, 543224, 543241, 543249, 543275, 543297, 543331, 543369, 543415, 543426, 543619, 543620, 543660, 543723, 543754, 543755, 543771, 543776, 543803, 543841, 544075, 544113, 544172, 544228, 544236, 544297, 544341, 544417, 544480, 544481, 544568, 544648, 544654, 544695, 544937, 544999, 545000, 545038, 545120, 545163, 545166, 545213, 545231, 545234, 545244, 545391, 545452, 545710, 545746, 545780, 545783, 545824, 545844, 545923, 545948, 545951, 546029, 546155, 546490, 546524, 546551, 546583, 546602, 547054, 547189, 547243, 547297, 547304, 547371, 547598, 547599, 547655, 547657, 547744, 547758, 547759, 547798, 547936, 547937, 548021, 548276, 548290, 548302, 548382, 548559, 548784, 549008, 549080, 549269, 549270, 549381, 549416, 549509, 549557, 549616, 549663, 550072, 550079, 550197, 550313, 550591, 550804, 550936, 550960, 551005, 551416, 551573, 551616, 551645, 551721, 551759, 551941, 551954, 551980, 552204, 552205, 552211, 552324, 552363, 552364, 552407, 552502, 552579, 552594, 552626, 552640, 552917, 552918, 552965, 552966, 553013, 553172, 553207, 553224, 553230, 553275, 553349, 553387, 553441, 553448, 553455, 553563, 553850, 553893, 553900, 553926, 554061, 554483, 554581, 554641, 554721, 554722, 554769, 555052, 555269, 555458, 555524, 555574, 555626, 555964, 555993, 556124, 556621, 556622, 556726, 556785, 556984, 557177, 557259, 557272, 558437, 558596, 558734, 558775, 41443799, 41444441, 41658234, 41862667, 42039161, and 537967 / 49371595.

Why it was recalled

Patient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked to the incorrect patient, which may adversely affect diagnosis and treatment.

Root cause (FDA determination)

Software design

Action the firm took

Philips Medical Systems issued URGENT-Device Correction letters dated January 27, 2009 to consignees describing the issue with the Easy Diagnost Eleva GXR RF systems. The consignees were informed of the potential for incorrect diagnosis or treatment if an incorrect image is used for diagnosis. Consignees should contact the Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FC0 70600030" with any communication.

Recalling firm

Firm
Philips Medical Systems North America Co. Phillips
Address
22100 Bothell Everett Hwy, Bothell, Washington 98021-8431

Distribution

Distribution pattern
Nationside Distribution: USA, including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PA, RI, SC, SD, TN, TX, TX, UT, VA, VT, WA, WI, and WV. Devices are not distributed to any foreign consignees.

Timeline

Recall initiated
2009-01-21
Posted by FDA
2010-11-26
Terminated
2010-12-30
Status

Source: openFDA Device Recall endpoint. Recall record ID #82093. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Product's marketing brochure is labeled in part: "Philips Easy Diagnost Eleva Conventional R/F syste · FDA Device Recalls