FDA Device Recalls

Recalls /

#82374

Product

Stryker 275 mL AutoFuser PainPump, single site continuous, 5 mL/hr, 2 mL bolus, 60 minute bolus refill, sterile, REF 0532-526-000.

FDA product code
MEBPump, Infusion, Elastomeric
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K041585, K060258
Affected lot / code info
Lots 08140754914 through 09119930611.

Why it was recalled

TFailure to infuse and potential compromised sterility: he seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.

Root cause (FDA determination)

Process control

Action the firm took

Stryker sent all consignees a Medical Device Recall Notification letter dated 5/14/09 and instructed to locate and destroy the recalled products and to complete and return the business reply card.

Recalling firm

Firm
Stryker Instruments Div. of Stryker Corporation
Address
4100 E. Milham Ave., Kalamazoo, Michigan 49001

Distribution

Distribution pattern
Nationwide and Canada.

Timeline

Recall initiated
2009-05-14
Posted by FDA
2009-09-04
Terminated
2010-01-20
Status

Source: openFDA Device Recall endpoint. Recall record ID #82374. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Stryker 275 mL AutoFuser PainPump, single site continuous, 5 mL/hr, 2 mL bolus, 60 minute bolus refi · FDA Device Recalls