Product

Zimmer Pulsavac Wound Debridement System high capacity intramedullary brush tip, sterile, Zimmer, Dover, Ohio; Catatog Number: 00515018300. Used in conjunction with the Pulsavac Plus Wound Debridement System for pulsatile lavage/debridement.

FDA product code
FQHLavage, Jet
Device class
Class 2
Medical specialty
General Hospital
Affected lot / code info
Lot Number: 61229259.

Why it was recalled

The seal on the sterile barrier pouch may be inadequate, resulting in lack of assurance of sterility.

Root cause (FDA determination)

Packaging process control

Action the firm took

Zimmer Orthopedic Surgical Products issued an "Urgent: Medical Device Recall" notice dated June 8, 2009 informing users of the affected devices and actions to be taken including the return of product and acknowledgement of receipt of notice by returning the Recall Certification Form by fax to 1-330-364-0974.

Recalling firm

Firm
Zimmer Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
Worldwide Distribution -- United States, Canada and Japan.

Timeline

Recall initiated
2009-06-08
Posted by FDA
2009-07-23
Terminated
2009-12-15