FDA Device Recalls

Recalls /

#83472

Product

Oxinium Femoral Head, 36 MM OD-3, 12/14 Taper, REF 71343603, Sterile, Smith & Nephew, Memphis TN 38116

FDA product code
JDIProsthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K030766
Affected lot / code info
Lot Numbers: 08BM09885, 08EM06396, 08FM14910, 08HM10480, 08KM07754, 08KM12526, 08HM03206, 08GM03007R, 08EM19444, 08EM19445, 08EM19446, 08EM20186, 08FM00231, 08RM00242, 08FM03526, 08FM04628, 08FM05689, 08FM06247, 08FM14906, 08FM17347, 08FM22085, 08FM22087, 08GM03007, 08GM03489, 08GM06660, 08GM11382, 08GM11425, 08GM12094, 08GM13061, 08GM17805, 08GM18609, 08HM02631, 08HM03926, 08HM10037, 08HM13107, 08HM13118, 08HM16147, 08HM16150, 08HM17621, 08HM18516, 08JM00370, 08JM04970, 08JM04971, 08JM05505, 08JM05506, 08JM08122, 08JM09041, 08JM11262, 08JM14571, 08JM15417, 08JM19246, 08JM19247, 08JM19248, 08JM20328, 08JM21716, 08JM21717, 08JM21718, 08KM03713, 08KM07732, 08KM14807, 08KM20680, 08KM23716, 08LM01328, 08LM02300, 08LM11759, 08LM12842, and 08LM13924

Why it was recalled

Radius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The deficiency may not be noticed by the user. Impingement and/or wear debris may lead to revision.

Root cause (FDA determination)

Other

Action the firm took

All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. Further questions are directed to the contacts at Smith and Nephew listed in the letter.

Recalling firm

Firm
Smith & Nephew Inc
Address
1450 E. Brooks Rd, Memphis, Tennessee 38116

Distribution

Distribution pattern
Nationwide; 17 international: Switzerland, Dubai, Brazil, Canada, South Africa, Australia, Belgium, Spain, Thailand Turkey, Costa Rica, Denmark, India, Italy, and France.

Timeline

Recall initiated
2009-01-23
Posted by FDA
2009-11-20
Terminated
2012-02-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #83472. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.