FDA Device Recalls

Recalls /

#84738

Product

Medtronic Intrathecal Catheter, model 8731SC, Method of Sterilization: Ethylene Oxide. Medtronic, Inc., Minneapolis, MN 55432-5604 USA. The 8731SC catheter is a two-piece catheter with a thin-walled spinal segment and a thicker-walled pump segment. These segments connect to each other with a connector pin. The pump segment connects to the pump with a sutureless pump connector. Contents: 38.1-cm spinal segment with length markings and guide wire 66-cm pump segment, separate connector pin, 15 T-guage introducer needles, Transparent strain-relief sleeves, opaque strain-relief sleeves, V-wing anchors. This is a component of an implantable drug infusion system, which consists of a drug infusion pump and a catheter. The drug infusion system stores and delivers parenteral drugs to the intrathecal space.

FDA product code
LKKPump, Infusion, Implanted, Programmable
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P860004S081
Affected lot / code info
Model 8731SC used with all IsoMed Pump Models including 8472. Not lot or serial number specific.

Why it was recalled

Current labeling for Medtronic Sutureless Connector (SC) catheters and revision kits (hereafter referred to as SC catheters) incorrectly states that SC catheters are compatible with Medtronic IsoMed constant-flow infusion pumps. SC catheters are compatible with Medtronic SynchroMed II and SynchroMed EL pumps. Medtronic has determined that SC catheters are not compatible with IsoMed pumps.

Root cause (FDA determination)

Device Design

Action the firm took

A "Medical Device Correction Notification" letter was issued on August 14, 2009 to implanting and managing physicians. The letter described the nature of the issue, scope of affected product, potential severity of the issue and provided recommendations. Physicians were instructed that Sutureless Connector (SC) catheters must not be used with IsoMed pumps for future revisions and implants. A copy of the letter was also sent to the Risk Manager at each hospital associated with physicians using SC catheters. Healthcare professionals were asked to complete and return the enclosed reply card. Direct questions to your Medtronic field representative or contact Medtronic Neuromodulation Technical Services at 1-800-707-0933. Patient Management information is also available at http://www.professional.medtronic.com under the Heading, "Advisories".

Recalling firm

Firm
Medtronic Neuromodulation
Address
7000 Central Ave NE, Minneapolis, Minnesota 55432-3568

Distribution

Distribution pattern
Worldwide Distribution -- US (including Washington D.C. and Puerto Rico), Algeria, Aruba, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Norway, Poland, Portugal, Reunion, Romania, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom and Vatican City.

Timeline

Recall initiated
2009-08-27
Posted by FDA
2009-09-10
Terminated
2011-12-12
Status

Source: openFDA Device Recall endpoint. Recall record ID #84738. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.