Recalls / —
—#85321
Product
Zimmer Natural Nail System, Antegrade Femoral interlock Module, tall, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-53. The device is used to drill into the proximal tibia or distal femur for placement of screws that stabilize the position of an intramedullary nail implant. The targeting guide instrumentation serves to assist the surgeon in precisely drilling into the bone and through holes that are located in the proximal and distal portion of the nail implant. After these holes are drilled, screws are placed into the holes to hold the implant in place in the bone.
- FDA product code
- HSB — Rod, Fixation, Intramedullary And Accessories
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K082770, K083497, K090596
- Affected lot / code info
- Lot 61221483.
Why it was recalled
The guides may come apart during use, some guides may experience surface cracking, guide handles and modules may not target the femoral nail holes accurately and the instruments may not be able to be assemble together due to a change in the threading method used.
Root cause (FDA determination)
Other
Action the firm took
On July 1, 2009, U.S. distributors were notified by e-mail to return two instruments. The recall was expanded with international accounts notified by e-mail on August 5, 2009, and U.S. accounts being notified by e-mail on August 6, 2009, to return 7 instruments. The scope of the recall was then expanded again to include 14 instruments and another letter dated September 14, 2009 to distributors and user accounts. The notifications described the affected products, problem, health risks and actions for consignees. Direct questions about the recall to Zimmer, Inc. by calling 1-800-613-6131.
Recalling firm
- Firm
- Zimmer Inc.
- Address
- 345 E Main St, Warsaw, Indiana 46580-2746
Distribution
- Distribution pattern
- Worldwide Distribution -- United States, France, Germany, Italy, Switzerland and United Kingdom.
Timeline
- Recall initiated
- 2009-07-01
- Posted by FDA
- 2009-11-24
- Terminated
- 2011-10-05
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #85321. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.