Product
10430054 Acuson with Physio Module, manufactured by Siemens Medical Solutions, Mountain View, CA.
- FDA product code
- ITX — Transducer, Ultrasonic, Diagnostic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K063803, K082142
- Affected lot / code info
- Model 10430054
Why it was recalled
ECG waveform does not accurately align with 2D, M-Mode or Spectral Images.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
The firm's Customer Safety Advisory notice was mailed to all affected customers. The letter describes when the malfunction occurs and the potential risk. It also recommended that users collect corresponding evidence with spectral waveforms that would confirm or deny findings based on ECG, in order to avoid potential risks for misdiagnosis. The firm planned to exchange the faulty hardware component in the system. Customers would be contacted to schedule for the hardware exchange. Questions regarding this issue should be directed to the local service support person.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc.
- Address
- 1230 Shorebird Way P.O. Box 7393, Mountain View, California 94043
Distribution
- Distribution pattern
- Worldwide Distribution
Timeline
- Recall initiated
- 2009-10-02
- Posted by FDA
- 2011-02-01
- Terminated
- 2011-03-29
- Status
- —