FDA Device Recalls

Recalls /

#87299

Product

Outlook Pump IV Set w/Universal Spike, Ultrasite Inj Site, B/C Valve and Pressure Limited Check Valve, 124 in US1502HP

FDA product code
FPASet, Administration, Intravascular
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K955585
Affected lot / code info
Product Code 470018. Lot numbers: 61048470 exp 2/28/2012, 61048471 exp 3/31/2012, 61048472 exp 4/30/2012

Why it was recalled

System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

The recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Questions are directed to the company's Customer Support Departments: US customers please call (800) 227-2862. Canadian customers, please call (800) 624-2920. International customers, please call (610) 691-5400.

Recalling firm

Firm
B. Braun Medical, Inc.
Address
901 Marcon Boulevard, Allentown, Pennsylvania 18109-9512

Distribution

Distribution pattern
The products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.

Timeline

Recall initiated
2009-11-24
Posted by FDA
2010-01-26
Terminated
2011-02-14
Status