FDA Device Recalls

Recalls /

#87303

Product

Outlook Pump IV Set w/Universal Spike, 1 Safeline & 3 Ultrasite Inj Sites, Check Valve, B/C Valve 119 in US1500XHP

FDA product code
FPASet, Administration, Intravascular
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K904518
Affected lot / code info
Product Code 470026. Lot numbers: 61048410 exp 3/31/2012, 61048411 exp 4/30/2012, and 61060357 exp 6/30/2012.

Why it was recalled

System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

The recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Questions are directed to the company's Customer Support Departments: US customers please call (800) 227-2862. Canadian customers, please call (800) 624-2920. International customers, please call (610) 691-5400.

Recalling firm

Firm
B. Braun Medical, Inc.
Address
901 Marcon Boulevard, Allentown, Pennsylvania 18109-9512

Distribution

Distribution pattern
The products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.

Timeline

Recall initiated
2009-11-24
Posted by FDA
2010-01-26
Terminated
2011-02-14
Status

Source: openFDA Device Recall endpoint. Recall record ID #87303. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.