FDA Device Recalls

Recalls /

#87953

Product

Medtronic HMS PLUS, Hemostasis Management System, Model 30514. It is a microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.

FDA product code
JOXAnalyzer, Heparin, Automated
Device class
Class 2
Medical specialty
Hematology
510(k) numbers
K894317
Affected lot / code info
Catalogue Number 30514

Why it was recalled

Recently the US Food and Drug Administration (FDA) issued an alert, communicating a change in the United States Pharmacopeia (USP) monograph for heparin, introducing a new USP reference standard that will in effect, change potency, and harmonize the USP unit dose with the WHO International Standard (IU) unit dose. Users of the Medtronic Hemostasis Management System (HMS Plus) must be aware o

Root cause (FDA determination)

Other

Action the firm took

Consignees were sent a Medtronic "Urgent Medical Device Notice' dated December 14, 2009. The letter was addressed to "Valued Customer". The letter described the product and the problem. Advised customers to configure the HMS Plus appropriately. The Field Notification was sent to all current HMS Plus users. The current HMS Plus users were identified by the sales of HMS Plus disposable product -Heparin Assay Cartridges from the period of August 1, 2008 to date. This time was considered appropriate, being more than twice the longest expiration date of any of the HMS disposable.

Recalling firm

Firm
Medtronic Inc
Address
710 Medtronic Pkwy, Minneapolis, Minnesota 55432-5603

Distribution

Distribution pattern
All states in USA (Including DC) except RI and VT.

Timeline

Recall initiated
2009-12-14
Posted by FDA
2010-03-11
Terminated
2011-12-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #87953. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.