Recalls / —
—#88035
Product
JOURNEY " UNI TIBIAL BASEPLATE, LEFT MEDIAL/RIGHT LATERAL, A/P 49 MM, M/L 29 MM, REF 71422224, FOR CEMENTED USE ONLY, QTY: (1), STERILE R, SMITH & NEPHEW, INC. MEMPHIS, TN 38116
- FDA product code
- HSX — Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K052265
- Affected lot / code info
- All Lot Numbers
Why it was recalled
The firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Root cause (FDA determination)
Device Design
Action the firm took
All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 12/23/2009. All affected Smith & Nephew International Distributors were notified via e-mail and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. For questions, please contact the firm, Leah Easley at 901-399-6137.
Recalling firm
- Firm
- Smith & Nephew Inc
- Address
- 1450 E. Brooks Rd, Memphis, Tennessee 38116
Distribution
- Distribution pattern
- US and Argentina, Switzerland, Australia, Canada, Belgium, Italy, Spain, Thailand, Austria; PROC, Finland, Germany, Malaysia, Portugal, Netherlands, Singapore, and England
Timeline
- Recall initiated
- 2009-01-23
- Posted by FDA
- 2010-03-03
- Terminated
- 2011-03-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #88035. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.