FDA Device Recalls

Recalls /

#88649

Product

SpeedLink 6.0 Transverse Connector Long 50-60mm, non-sterile, 1304-60L

FDA product code
KWPAppliance, Fixation, Spinal Interlaminal
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K002082
Affected lot / code info
10ZG, 12LY, 13XB, 15DZ, 30HP, 36FA, 40EG, 45XB, 7EB.

Why it was recalled

The firm received complaints that the device was damaged or fractured while the surgeon was bending it prior to implantation. There were also complaints that the cam may dislodge during tightening. The firm added additional instructions for use to reduce the probability of these two types of events.

Root cause (FDA determination)

Other

Action the firm took

The firm provided a revised surgical technique document to the Zimmer spine sales force in January, 2009. The firm will provide an Urgent Medical Device Correction notification to speedlink users through the sales force.

Recalling firm

Firm
Zimmer Inc.
Address
345 E Main St, Warsaw, Indiana 46580-2746

Distribution

Distribution pattern
The devices are contained in implant cases stocked by Zimmer spine and made available to the Zimmer sales force for surgeries. They are not in general distribution.

Timeline

Recall initiated
2009-01-01
Posted by FDA
2010-10-08
Terminated
2011-01-05
Status