Product
PrecisePLAN 2.10 Treatment Planning System
- FDA product code
- MUJ — System, Planning, Radiation Therapy Treatment
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K022411
- Affected lot / code info
- 1762685762, 1762688017, 1762688506, 1726288507, 1762676911, 1762688175, 1762689384, 1762688019, 1762688022, 1762688876, 1762689442, 1762688177, 1762690813, 1762686657, 1762689588, 1762689386, 1762684357, 1762686833, 1762684370, 1762678238, 1762689519, 1762684359, 1762690814, 1762688514, 1762578223, 1762678239, 1762684360, 1762684361, 1762684362, 1762684367, 1762683737, 1762682414, 1762688081, 1762686584, 1762686843, 1762686832, 1762688024, 1762688080, 1762686836, 1762684358
Why it was recalled
Error in dose rates which were calculated at exactly twice the expected values.
Root cause (FDA determination)
Software design
Action the firm took
As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.
Recalling firm
- Firm
- Elekta, Inc.
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, Georgia 30092-3011
Distribution
- Distribution pattern
- Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland
Timeline
- Recall initiated
- 2006-03-15
- Posted by FDA
- 2010-04-16
- Terminated
- 2010-12-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #89893. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.