Recalls / —
—#89894
Product
Elekta Synergy XVI Medical Linear Accelerator. Model Numbers R3.5, R4.0 and R4.2. Elekta, Inc. Norcross, GA 30092. Intended for use as a radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- FDA product code
- IYE — Accelerator, Linear, Medical
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K051932
- Affected lot / code info
- Serial Numbers: 151008, 151024, 151025, 151033, 151038, 151051, 151053, 151055, 151059, 151073, 151074, 151076, 151080, 151084, 151093, 151104, 151110, 151130, 151134, 151141, 151143, 151151, 151154, 151155, 151156, 151157, 151160, 151167, 151168, 151173, 151176, 151178, 151216, 151228, 151229, 151238, 151245, 151250, 151253, 151256, 151259, 151260, 151266, 151267, 151269, 151298, 151301, 151305, 151309, 151317, 151325, 151326, 151327, 151328, 151340, 151352, 151359, 151360, 151362, 151363, 151369, 151371, 151387, 151394, 151398, 151411, 151416, 151421, 151422, 151423, 151430, 151433, 151435, 151441, 151446 and 151479.
Why it was recalled
It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.
Root cause (FDA determination)
Software design
Action the firm took
Elekta, Inc. issued an "Important Notice" dated December 12, 2008. Customers were informed of the affected device and instructed on proper steps until an update is provided. For further information, contact Elekta, Inc. at 1-770-300-9725.
Recalling firm
- Firm
- Elekta, Inc.
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, Georgia 30092-3011
Distribution
- Distribution pattern
- Nationwide Distribution -- AL, AZ, CO, CT, FL, GA, ID, KS, KY, MA, MI, MN, MS, NE, NY, OH, OK, OR, PA, TN, TX, VA, WA and WV.
Timeline
- Recall initiated
- 2008-12-12
- Posted by FDA
- 2010-04-16
- Terminated
- 2010-12-28
- Status
- —