FDA Device Recalls

Recalls /

#90458

Product

Tubing Assembly #142, V2, V12 Mixer/Wash a component of the DxC instruments. Tube, Assembly, #142 V2, V12 Mixer/Wash delivers appropriate amount of wash concentrate on the UniCel DxC SYNCHRON Clinical Systems. The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent)

FDA product code
JJEAnalyzer, Chemistry (Photometric, Discrete), For Clinical Use
Device class
Class 1
Medical specialty
Clinical Chemistry
510(k) numbers
K042291, K060256
Affected lot / code info
All serial numbers

Why it was recalled

The recall was initiated because the Wash Concentrate Tube Assembly 142 may fail to deliver the appropriate amount of wash concentrate on some SYNCHRON DxC Systems due to an accumulation of crystals. The impact of this issue would include: (1) Quality Control and patient results can potentially be affected. (2) If system has been calibrating properly and Quality Control is recovering within es

Root cause (FDA determination)

Pending

Action the firm took

Beckman Coulter sent out an "Urgent Product Corrective Action" letter dated July 23, 2009 and mailed the week of July 27, 2009 to all affected customers. The letter provides the cusotmers with explanation of the problem identifed and asks them to take the following actions: " If consignees are not able to resolve these issues through routine troubleshooting they were told to contact BCI Customer Technical Support at (800) 854-3633 in the United States and Canada. " This issue will be corrected with an upcoming software version and tubing modification. " Complete and return the enclosed response form within 10 days so that Beckman Coulter may maintain their records. Consignees were also asked to share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. Questions regarding this Product Corrective Action letter were directed to their Customer Technical Support Center (Hotline) at (800) 854-3633 in the United States and Canada.

Recalling firm

Firm
Beckman Coulter Inc.
Address
200 S Kraemer Blvd, Brea, California 92821-6208

Distribution

Distribution pattern
Worldwide Distribution - USA (Nationwide) and Canada

Timeline

Recall initiated
2009-07-23
Posted by FDA
2011-06-23
Terminated
2012-07-02
Status

Source: openFDA Device Recall endpoint. Recall record ID #90458. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.