Product
Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00, sterile, Zimmer Inc., Parsippany, NJ. Usage: Implantable devices that are used to address bony deficiencies of the pelvic antebulum in revision total hip arthroplasty surgery.
- FDA product code
- LPH — Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K061067
- Affected lot / code info
- 60492196, 60531086, 60703467, 60667441, 60728123, 60767106, 60787931, 60808260, 60904605, 61024439, 61143012, 61217116 and 61355988.
Why it was recalled
The firm has determined that some of the packages could develop a breach of the seal in the sterile barrier pouches.
Root cause (FDA determination)
Packaging
Action the firm took
The firm sent an "Urgent Device Removal" letter dated 4/6/2010. The consignees were instructed to inventory the products on hand and return the recalled product to the manufacturer. More information is available at 1-800-613-6131.
Recalling firm
- Firm
- Zimmer Inc.
- Address
- 345 E Main St, Warsaw, Indiana 46580-2746
Distribution
- Distribution pattern
- US, Germany, Brazil, Singapore, Canada, Australia, England, India, and Switzerland.
Timeline
- Recall initiated
- 2010-04-06
- Posted by FDA
- 2010-08-12
- Terminated
- 2011-10-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #90688. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.