Product

Howmedica Osteonics Corp., Stryker Orthopaedics Passport A.R. Knee Instrumentation Anterior Skim Guide; Catalog Number 7650-5003.

FDA product code
LXHOrthopedic Manual Surgical Instrument
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
Catalog Number 7650-5003; all lots.

Why it was recalled

One of the two posts on the Anterior Resection Guide could disassociate from the main body of the part which has been determined to be caused by a fracture weld.

Root cause (FDA determination)

Device Design

Action the firm took

Stryker Urgent Product Recall letters dated April 5, 2010, were sent out via Federal Express to Stryker branches/agencies, Hospital Risk Management, hospital Chief of Orthopedics and Surgeons. All hospitals and surgeons that used the product in the last year were notified. Questions and/or concerns should be directed to Rich Wolyn at 201-972-2100.

Recalling firm

Firm
Stryker Howmedica Osteonics Corp.
Address
325 Corporate Drive, Mahwah, New Jersey 07430

Distribution

Distribution pattern
Worldwide distribution: USA, Ireland, Germany, Sweden, The Netherlands, France, Italy, Switzerland, South Africa, United Kingdom, Japan, Argentina, Mexico, Miramar (city), China, Taiwan, India, Australia, Korea, New Zealand, Malaysia, Singapore, and Canada.

Timeline

Recall initiated
2010-04-05
Posted by FDA
2010-08-02
Terminated
2012-08-28