Recalls / —
—#90708
Product
Howmedica Osteonics Corp., Stryker Orthopaedics Passport A.R. Knee Instrumentation Anterior Skim Guide; Catalog Number 7650-5003.
- FDA product code
- LXH — Orthopedic Manual Surgical Instrument
- Device class
- Class 1
- Medical specialty
- Orthopedic
- Affected lot / code info
- Catalog Number 7650-5003; all lots.
Why it was recalled
One of the two posts on the Anterior Resection Guide could disassociate from the main body of the part which has been determined to be caused by a fracture weld.
Root cause (FDA determination)
Device Design
Action the firm took
Stryker Urgent Product Recall letters dated April 5, 2010, were sent out via Federal Express to Stryker branches/agencies, Hospital Risk Management, hospital Chief of Orthopedics and Surgeons. All hospitals and surgeons that used the product in the last year were notified. Questions and/or concerns should be directed to Rich Wolyn at 201-972-2100.
Recalling firm
- Firm
- Stryker Howmedica Osteonics Corp.
- Address
- 325 Corporate Drive, Mahwah, New Jersey 07430
Distribution
- Distribution pattern
- Worldwide distribution: USA, Ireland, Germany, Sweden, The Netherlands, France, Italy, Switzerland, South Africa, United Kingdom, Japan, Argentina, Mexico, Miramar (city), China, Taiwan, India, Australia, Korea, New Zealand, Malaysia, Singapore, and Canada.
Timeline
- Recall initiated
- 2010-04-05
- Posted by FDA
- 2010-08-02
- Terminated
- 2012-08-28
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #90708. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.