Product

Arrow 4FR Peripherally Inserted Central Catheter (PICCs) Kits. Product numbers: ASK-01451-BMH; ASK-05041-CHC; ASK-05041-CHC1; ASK-05041-CHM; ASK-05041-HMC; ASK-05041-QV; ASK-07041-BMH; CN-05041-LW; MSO-01451-UCL; MTO-01451-RH; PL-05041; PR-04041-HPX; PR-05041; PR-05041-HP; PR- 05041-HPX; PR-05041-LW; PR-05041-MW; PR-05041-T; PR-05042; PR-05541-HPX; PR-07041-PTSP; and RJ-01451-W.

FDA product code
LJSCatheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K061289, K073451
Affected lot / code info
Product number ASK-01451-BMH - Lot numbers RF8060214 and RF8096394; Product number ASK-05041-CHC - Lot numbers RF8109279 and RF9056727; Product number ASK-05041-CHC1 - Lot numbers RF9069677 and RF9098408; Product number ASK-05041-CHM - Lot numbers RF8057827 and RF8068899; Product number ASK-05041-HMC - Lot numbers RF8068662, RF8083994, RF8108545, RF9043625, RF9056403, RF9085278, and RF9098580; Product number ASK-05041-QV - Lot number RF9097395; Product number ASK-07041-BMH - Lot numbers RF9031839, RF9098784, and RL8108648; Product number CN-05041-LW - Lot numbers RF8069608, RF9056127, RF9070920, and RL9019402; Product number MSO-01451-UCL - Lot numbers RF9057410, RF9084460, RL8118758, and RL9029477; Product number MTO-01451-RH - Lot number RL9019228; Product number PL-05041 - Lot numbers RF8071766 and RF9109677; Product number PR-04041-HPX - Lot numbers RF8072920 and RF8109646; Product number PR-05041 - Lot numbers RF8056742, RF8068423, RF8069651, RF8071677, RF8095208, RF8095832, RF8108465, RF9028902, RF9043449, RF9056132, RF9085152, and RF9086357; Product number PR-05041-HP - Lot numbers RF8072840, RF8083506, RF8084926, and RF8107907; Product number PR-05041-HPX - Lot numbers RF8071099, RF8072981, RF9015967, RF9057607, and RF9097057; Product number PR-05041-LW - Lot numbers RF8057252, RF8072743, RF8083529, RF8108466, RF8110585, RF9031472, RF9044054, RF9060502, and RF9097058; Product number PR-05041-MW - Lot numbers 8057253, RF8060216, RF8095834, RF8108467, RF8110586, RF9016412, RF9031473, RF9044055, RF9069045, RF9071244, RF9085152, RF9086354, RF9100412, and RF9109679; Product number PR-05041-T - Lot numbers RF8068845 and RF8107962; Product number PR-05042 - Lot numbers RF8068846, RF8107963, RF8122680, RF9016413, RF9044057, and RF9073297; Product number PR-05541-HPX - Lot number RF8124014; Product number PR-07041-PTSP - Lot numbers RF8068850, RF8071679, RF8107965, RF8110358, RF9042708, and RF9072922; and Product number RJ-01451-W - Lot numbers RF8068664, RF8096774, RF8123596, and RF9060298.

Why it was recalled

Tight fit in sheath. Arrow has received complaints of the PICC catheter fitting too tightly in the "kit supplied" peelable sheath. In some cases, clinicians have been unable to insert the catheter through the peelable sheath during the insertion procedure. If the catheter will not pass through the sheath, the clinician is required to place a guide wire back into the sheath, remove the sheath, a

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Arrow International issued an Urgent Medical Device Recall letter dated 3/29/10 to inform their customers of the problem and the need to return the product.

Recalling firm

Firm
Arrow International Inc
Address
2400 Bernville Road, Reading, Pennsylvania 19605

Distribution

Distribution pattern
Worldwide distribution: USA, Australia, Brazil, Canada, New Zealand, Hong Kong, Malaysia, and Chile.

Timeline

Recall initiated
2010-03-29
Posted by FDA
2010-05-26
Terminated
2012-09-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #90745. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.