Recalls / —
—#91339
Product
The Dimension NTP and LNTP Flex reagent cartridge (N- Terminal Pro- Brain Natriuretic Peptide and N-Terminal Pro brain Natriuretic Peptide (low volume)
- FDA product code
- NBC — Test, Natriuretic Peptide
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K082645
- Affected lot / code info
- Lot# FD0110 Exp# 04/20/2010, Lot#EB0122, EXP# 05/02/2010, Lot# FC0190, EXP# 07/09/2010, Lot# FC0219, EXP# 08/07/2010, Lot# FB0256, EXP#09/13/2010, Lot# FC0100, EXP#04/10/2010, Lot#EA0122, EXP#02/05/2010, Lot# FB0190, EXP# 07/09/2010, Lot# FB0219, EXP# 08/07/2010, lot# FA0256, EXP# 09/13/2010.
Why it was recalled
Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Root cause (FDA determination)
Other
Action the firm took
Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 500 GBC Drive PO BOX 6101, Mailstop 514, Newark, Delaware 19714-6101
Distribution
- Distribution pattern
- Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.
Timeline
- Recall initiated
- 2010-04-12
- Posted by FDA
- 2010-11-24
- Terminated
- 2013-03-19
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #91339. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.