—#92243
Product
Stryker Bur Guard REF 2296-301, Stryker Instruments, Kalamazoo, MI (USA) The Impaction Bur Guard is intended to be used with the Impaction Drill. The drill is used for surgical procedures involving precision drilling and cutting of bone and hard tissue. The bur guard covers the nose of the handpiece and the shaft of the cutting accessory to prevent exposure to heat and moving parts, as well as minimize bur whip and handpiece vibration.
- FDA product code
- HWE — Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
- Device class
- Class 1
- Medical specialty
- General, Plastic Surgery
- 510(k) numbers
- K961970
- Affected lot / code info
- Replace by dates of May 2009 through August 2010
Why it was recalled
The reason for the field action is that the inner diameter of the Bur Guards may be out of specification. If there is excess material on the inside of the Bur Guard at the location where the guard tapers, it may come into contact with the Nose Cone of the drill. This could result in an injury or burn to the patient and/or user.
Root cause (FDA determination)
Device Design
Action the firm took
Stryker Instruments sent an Urgent Medical Device Recall Notification dated June 10, 2010, to their customers. The letter described the product the problem and the action to be taken by the customers. The customers were instructed to: 1) immediately review this recall notification. 2) forward this letter and the attached Business Reply Form (BRF) to all affected locations. 3) Immediately check and quarantine any affected product found. 4) Destroy all affected product by breaking off at least 3 of the tabs on the bur guard and then dispose of the affected product in normal trash. 5) Complete and sign the BRF and fax to Stryker Instruments Regulatory Department, 866-521-2762. NOTE: Upon receipt of the BRF, a replacement bur guard will be sent to the customer's account for each affected bur guard destroyed. For questions regarding this recall please contact Stryker Instruments: 1-800-800-4236, ext. 3808 jennifer.mars@stryker.com
Recalling firm
- Firm
- Stryker Instruments Div. of Stryker Corporation
- Address
- 4100 E. Milham Ave., Kalamazoo, Michigan 49001
Distribution
- Distribution pattern
- Worldwide Distribution: US AND COUNTRIES OF SCANDINAVIA, AUSTRALIA, CANADA, CHILE, CHINA, BRAZIL, FRANCE, GERMANY, GREECE, SOUTH AFRICA, ITALY, JAPAN, MEXICO, SWITZERLAND, PORTUGAL, SINGAPORE and UK.
Timeline
- Recall initiated
- 2010-06-02
- Posted by FDA
- 2010-09-13
- Terminated
- 2011-04-21
- Status
- —